CAEN
Electronic Instrumentation
UM2619 – N1419 User Manual
29
Remote Control: format EEPROM
By typing
F
it is possible to access the format EEPROM menu:
Fig. 52: Format EEPROM Menu
After the FORMAT command, all the channels have the following settings:
Vset = 0 V
Iset = 21.0 µA
Ramp Up / Down = 5 V/s
Trip = 10 s
MaxV = 510 V
Power Down = Kill
Module setting:
Interlock Mode = Active CLOSED
Remote Control: Current offset calibration
By typing
O
it is possible to perform the current offset calibration (the calibration must be performed after
30minutes operation at least); it is necessary to perform a calibration anytime external operating conditions
change; the following instructions are displayed:
Fig. 53: Current offset calibration warning
Once channels are off and disconnected from the load, type
y
; at this point the program asks for the current value
(VSET) selected for the offset compensation, such value is common to all channels; factory calibration is performed
at 250V, type VSET then press <ENTER>
Fig. 54: Current offset calibration VSET selection