CAEN
Electronic Instrumentation
UM2619 – N1419 User Manual
21
Fig. 29: Channel KILL screen
In order to access another channel, the EXIT parameter has to be selected
Fig. 30: Channel EXIT screen
Now by turning the TUNE ROTARY SWITCH another channel number to be set can be selected.
If CONTROL MODE (see p.24) is set to REMOTE, the left lower row reports DIS (Disabled), since the channel can be
accessed only via the serial links (see p.24). If the INTERLOCK MODE is changed while one channel is ON, the
channel is turned OFF and the left lower row reports ILK (Interlock); if the channel is OFF, it cannot be turned ON,
until it is enabled according to the Interlock logic (see p.11).
Group Settings
5
Group settings allow to broadcast the same parameter value to all channels.
In order to operate Group settings:
Turn the TUNE ROTARY SWITCH until ALL is displayed in the left column
Fig. 31: Group selection
Push the TUNE ROTARY SWITCH: at this point the frame of the left column becomes red and the GROUP is selected.
Turn the TUNE ROTARY SWITCH until the parameter to be set (for example VSET) is displayed in the right column (all
four channels values).
Fig. 32: Group active
5
Mod. N1419B has not group settings; Mod. N1419A has group settings, 2 channels values are displayed