CAEN
Electronic Instrumentation
UM2619 – N1419 User Manual
14
Technical specifications table
Table 4: Mod. N1419 Channel technical specifications
Output channels:
Positive or Negative Polarity (requires internal setting, see p.
Errore. Il segnalibro non è definito.
)
Output ranges:
500 V / 200µA
Max. Ch. Output Power:
100 mW
Vset / Vmon Resolution:
10 mV
Iset / Imon Resolution
2
:
If IMON RANGE = High is selected resolution is 5 nA
If IMON RANGE = Low is selected resolution is 0.5 nA
Vmax:
0 ÷ 510 V
Absolute maximum HV level that the channel is allowed to reach, independently from the preset value
Vset. Output voltage cannot exceed the preset value Vmax. The accuracy is 1 % ± 0.5 V
Vmax resolution:
± 0.1 V
Alarm output:
Open collector, 100 mA maximum sink current
Interlock input:
LOW: <1V; current~5mA; HIGH: 4÷6 V
Ramp Up/Down:
1÷50 Volt/s, 1 Volt/s step
Trip:
Max. time an "overcurrent" is allowed to last (seconds). A channel in "overcurrent" works as a current
generator; output voltage varies in order to keep the output current lower than the programmed
value. "Overcurrent" lasting more than set value (1 to 9999) causes the channel to "trip". Output
voltage will drop to zero either at the Ramp-down rate or at the fastest available rate, depending on
Power Down setting; in both cases the channel is put in the OFF state. If trip= INFINITE, "overcurrent"
lasts indefinitely.
TRIP range: 0 ÷ 999.9 s; 1000 s = Infinite. Step = 0.1 s
Vmon vs. Vout Accuracy:
3
±0.02% of read value ±0.2V
Vset vs. Vmon Accuracy:
4
±0.02% of read value ±0.2V
Imon vs. Iout Accuracy:
4
If IMON RANGE = High : ±2% of read value ±20nA
If IMON RANGE = Low: ±2% of read value ±2nA
Iset vs. Imon Accuracy:
4
If IMON RANGE = High : ±2% of read value ±30nA
If IMON RANGE = Low: ±2% of read value ±3nA
Voltage Ripple:
4
band width
typical
maximum
10Hz÷100Hz
<5mVpp
<5mVpp
100Hz÷100MHz
<5mVpp
<5mVpp
Humidity range:
0 ÷ 80%
Operating temperature:
0 ÷ 45°C
Storage temperature:
-10 ÷ 70°C
Vout / Temperature coefficient:
max. 50ppm / °C
Vout /voltage coefficient:
max 2ppm/mV
Imon / Temperature coefficient:
max 100ppm/C°; max 300ppm/C° with Imon X10 zoom (optional)
Long term stability Vout vs. Vset:
± 0.02% (after one week @ constant temperature)
2
The module is calibrated by introducing a positive offset on the current monitor, see details p. 24
3
From 10% to 90% of Full Scale Range
4
Measured with: 1m cable length; 2nF capacitance