12
Table 3: Interlock operation
CONFIGURATION
↓
INTERLOCK MODE (p.15)
→
OPEN
CLOSE
leave contact open
INTERLOCK
ENABLED
voltage level (0÷1V, ~5mA current) between pin 2 and pin 3
INTERLOCK
ENABLED
short circuit pin 1 with pin 2, and pin 3 with pin 4
ENABLED
INTERLOCK
voltage level (4÷6V, ~5mA current) between pin 2 and pin 3
ENABLED
INTERLOCK
The front panel Interlock LED is ON when the INTERLOCK is enabled; as INTERLOCK is enabled, channels are turned
off at the fastest available rate, regardless the RAMP DOWN setting.
Remote communication control section
Fig. 11: Remote communication control and RS485 I/O – RS232 IN electrical scheme
NAME:
TYPE:
FUNCTION:
IN
AMP 280371-2
RS485 Input
1
; adaptable to RS232 standard
OUT
AMP 280371-2
RS485 Output
USB
B TYPE USB
USB2.0 compliant
realized via USB ↔ RS232 FT232BM converter
1
RS 485 Serial Port Interface allows to control up to 32 modules connected by a twisted pair cable; the first and last modules must be terminated,