803x Series 8-16 Ch. Programmable HV Power Supplies
16
N.B. parameters written in
italic
are available only on x8031 series.
For each channel the following parameters can be programmed and monitored either locally or remotely:
Parameter:
Function:
Unit:
(±)
Channel polarity
Vmon
High Voltage Monitored value
Volt
Imon
Current Monitored value
µA
Status
ON/OFF; Ramp UP/DOWN; OVV; UNV; OVC; OVP; MAXV; TRIP;
OVT; OFF; KILL; ILK; CAL_ERR
Vset
High Voltage programmed value
Volt
Iset
Current Limit programmed value
µA
MaxV
Absolute maximum High Voltage level that the channel can reach
(see p. 12)
V
Ramp-Up
Maximum High Voltage increase rate
V/s
Ramp-Down
Maximum High Voltage decrease rate
V/s
Power Down
Power Down mode after channel TRIP
KILL or RAMP
Trip
Max time "overcurrent" allowed to last (1000
= ∞)
s
Imon Range
Current Monitor Zoom
H or L
Parameters available only with x8031 series:
VTC
Enable Temperature compensation
KT
Coefficient of Temperature compensation
V/°C
T Ref
Reference Temperature
°C
T Step
Programmed Interval of Temperature compensation
°C
T Ext
External Temperature
°C
To set one parameter, tap on the relevant name, and change and/or enter the desired value through the
“virtual keypad” (see below); confirm with “Enter”.
Tap the red arrow to go back to Main Menu.
1
2
3
4
5
6
7
8
9
.
0
Del
Enter
Group Settings
This menu allows to extend settings to all the channels of the unit, and to monitor all of them.
GROUP MENU
←
Ch
0(-)
1(-)
2(-)
3(-)
VMon
IMon
Status
VSet
Zoom
ISet
Chan
Page
1/3