803x Series 8-16 Ch. Programmable HV Power Supplies 

11 

Remote communication control section 

 

USB 

B TYPE USB 

USB2.0 compliant  

ETH 

10/100 Base-T female connector 

 

HV Channel Output 

 

Channel Output:

 

8031 series

: BNC RADIALL R141.557.000 

Electrical: 

Impedance  50Ω

;  Operating  temp.  -65/+165°C; 

Frequency 0-4 GHz; VSWR 1.06+0.006 xF(GHz) Max; Insertion loss 

.025*√F(G

Hz) dB Max; RF leakage 

 (57 - F(GHz)) dB Max; Voltage 

rating  500Veff  Max  ;  Dielectric  withstanding  voltage  1500Veff 
min; Insulation resistance 

5000MΩ

 min; 

Mechanical: Center contact retention: Axial force 

 Mating  end 

18 N min; Axial force 

 Opposite end 18N min; Torque 2.8 N.cm 

min Recommended torque: Panel nut 250 N.cm; Mating life 500 
Cycles min Weight 8.000 g

 

Channel Output:

 

8032-33-34-34H series

: SHV RADIALL R317580; 

Impedance: 50 Ohm; Frequency range: 0 

 

2 GHz; VSWR: <1.20 + 0.3 F (GHz) 

 (plug and jack); Test voltage: 

10kV DC 

 1mn (unmated 

connectors);  Ratings:  12kV  DC 

  1mn  (mated  pairs);  Current 

rating: 10 A  

MON 

AMP  280371-2;  The  test  points  allow  to  monitor  the 

Channel Output Voltage, Current and Temperature. 
V/I analog monitor output; range: 0 ÷ +5V; proportional to output 
voltage and current.  
T: input for PT1000 temperature probe (8031 series only). 

 

WARNING! These connectors produce extremely hazardous high voltages at a potentially lethal 
current  level;  never  connect  or  disconnect  the  HV  OUT  connector  with  the  power  ON/OFF 
switch ON; always switch power OFF and wait at least 30s before connecting or disconnecting 
HV cables. 

AC Input (DT803x, R803x) 

 

IEC 60 320 Socket with switch; to be connected to Mains 100 - 240 Vac (50 - 60 Hz) via provided power 
cord.  
Two Fuses: 6.3x32 (8 channels: 1A; 16 channels: 2A); Retarded 250VAC 
 

Содержание 803 Series

Страница 1: ...803x Series 8 16 Ch Programmable HV Power Supplies Rev 10 19 December 2022...

Страница 2: ...ible for damage caused by improper use of the device The manufacturer declines all responsibility for damage resulting from failure to comply with the instructions for use of the product The equipment...

Страница 3: ...er ON 13 Local Control 14 BOARD Settings 14 Channel settings 15 Group Settings 16 Status Icon 18 Remote Control 19 Software tools 19 GECO2020 19 CAEN HV Wrapper 19 Power Supply Modules LabVIEW Instrum...

Страница 4: ...ernating Current The following symbol may be reported in the present manual General warning statement The symbol could be followed by the following terms DANGER indicates a hazardous situation which i...

Страница 5: ...ID ELECTRIC SHOCK OR FIRE HAZARD DO NOT OPERATE THIS PRODUCT WITH COVERS OR PANELS REMOVED CAUTION Do Not Operate in Wet Damp Conditions TO AVOID ELECTRIC SHOCK DO NOT OPERATE THIS PRODUCT IN WET OR D...

Страница 6: ...ivered through BNC connectors other versions feature SHV s The HV output RAMP UP and RAMP DOWN rates may be selected independently for each channel Safety features include OVERVOLTAGE and UNDERVOLTAGE...

Страница 7: ...Channels Voltage 100 240V Frequency 50 60Hz 0 8A rms max Fuse 2xT1A 6 3x32 250VAC Voltage 100 240V Frequency 50 60Hz 1 6A rms max Fuse 2xT2A 6 3x32 250VAC Voltage 100 240V Frequency 50 60Hz 1 6A rms m...

Страница 8: ...CAEN Electronic Instrumentation 803x Series 8 16 Ch Programmable HV Power Supplies 8 Front panel...

Страница 9: ...AEN Electronic Instrumentation 803x Series 8 16 Ch Programmable HV Power Supplies 9 External connections Local control section 2 8 LED Touch Screen Parameter and Mode setting Local settings monitoring...

Страница 10: ...ferences can be used to provide a TTL compatible level on pin 2 and 3 In the first case externally supplied device the maximum allowed ratings are Maximum voltage between and 12V Maximum sink current...

Страница 11: ...nnel Output 8032 33 34 34H series SHV RADIALL R317580 Impedance 50 Ohm Frequency range 0 2 GHz VSWR 1 20 0 3 F GHz plug and jack Test voltage 10kV DC 1mn unmated connectors Ratings 12kV DC 1mn mated p...

Страница 12: ...he channel to trip Output voltage will drop to zero either at the Ramp down rate or at the fastest available rate depending on Power Down setting in both cases the channel is put in the off state If t...

Страница 13: ...M mechanics The unit is an equipment for BUILDING IN install it in a NIM crate providing the required power supplies with forced cooling air flow then connect the HV Channel Output to the load R803x s...

Страница 14: ...itor values are available also with remote control include Parameter Type Function Status Monitor Module power supply status HVMax Monitor Hardware Max Voltage settable via trimmer see p 10 HIMax Moni...

Страница 15: ...left arrow to go back The right arrow allows to access to DHCP settings MAC Address 00 12 5e 00 07 07 DHCP Enabled If a DHCP Server is available then the module can be enabled or disabled as DHCP cli...

Страница 16: ...ate V s Ramp Down Maximum High Voltage decrease rate V s Power Down Power Down mode after channel TRIP KILL or RAMP Trip Max time overcurrent allowed to last 1000 s Imon Range Current Monitor Zoom H o...

Страница 17: ...VC OVP MAXV TRIP OVT OFF KILL ILK CAL_ERR single channels Vset High Voltage programmed value Volt Iset Current Limit programmed value A MaxV Absolute maximum High Voltage level that the channel can re...

Страница 18: ...CAEN Electronic Instrumentation 803x Series 8 16 Ch Programmable HV Power Supplies 18 Status Icon Three types of Icon in the display status area indicate OK status WARNING status ALARM status...

Страница 19: ...r software libraries For more info please visit www caen it products firmware software section Power Supply Modules LabVIEW Instrument Driver Power Supply Modules LabVIEW Instrument Driver is a set of...

Страница 20: ...it 1 Flow control none Launch communication Type caen then enter Ethernet Launch the terminal emulator select port number 1470 Please note that line editing must be disabled prior to Ethernet access E...

Страница 21: ...e to access the format EEPROM menu Channels settings By typing D on Terminal Board Menu it is possible to access channels settings To change one parameter point the parameter with the arrow keys and t...

Страница 22: ...tation 803x Series 8 16 Ch Programmable HV Power Supplies 22 By typing G accesses Group Mode settings are extended to all channels I changes Interlock Mode see p 10 R resets alarm conditions see p 10...

Страница 23: ...es 25 8 C Vout 50 4V If Text becomes 25 9 C Vout 50 4V If Text becomes 26 C Vout 50 5V The compensation is signed in fact if temperature decreases voltage is reduced If Text becomes 24 Vout 49 5V By i...

Страница 24: ...se to a command of MON or INFO type it is not used in response to a command of SET type successful response to a command of SET type CMD OK CR LF successful response to a command of MON type CMD OK VA...

Страница 25: ...efine for parameter type described by data format define NUMERIC 0 define ONOFF 1 define CHSTATUS 2 define BDSTATUS 3 define BINARY 4 define STRING 5 define ENUM 6 define READ 0 define WRITE 1 define...

Страница 26: ...ON EN and one of these conditions is TRUE Bit 15 MAXV 1 VMON HVMAX set via trimmer Power On in local mode Switch from remote to local TRIP TRIP time value VTC Enable Temperature compensation Only for...

Страница 27: ...s or flammable liquids on the screen Cleaning the air vents It is recommended to occasionally clean the air vents if present on all vented sides of the board Lint dust and other foreign matter can blo...

Страница 28: ...ervice it is recommended to perform the following actions Detach all the signal input output cable Wrap the device in its protective packaging Insert the device in its packaging if present THE DEVICE...

Страница 29: ...A Via Vetraia 11 55049 Viareggio Italy Phone 39 0584 388 398 Fax 39 0584 388 959 info caen it www caen it CAENspa INDIA Private Limited B205 BLDG42 B Wing Azad Nagar Sangam CHS Mhada Layout Azad Naga...

Отзывы: