EN / 12
7.1 Cleaning
Before cleaning, the optical 3D scanner must be put out of operation. To do so, unplug the scanner. Use only dry cloths for cleaning. The
scanner may be vacuumed or blown out. However, pay attention to the optical elements so that they do not become misaligned (camera
focus). To prevent damage to the optical sensors located in the upper part of the scanner, they may only be dusted gently using a microfi -
ber cloth.
NOTE!
The optical sensor system in the top part of the scanner is extremely sensitive.
DANGER!
Accidentally turning on the scanner during cleaning.
Do not use water, steam or cleaning agents for cleaning.
7.2 Recalibration
To guarantee a consistent scan result, it is recommended to use the supplied
calibration plate to perform a calibration once weekly. The scanner should also be calibrated after transport or if it is moved to a different
location. Instructions for recalibrating your scanner can be found in our software manual. Start the scanning process using software and
proceed step by step according to the software instructions. If anything is unclear, contact our customer service department (see chapter
"Customer service").
Make sure that the circles on the glass plate are not damaged, otherwise calibration is no longer possible.
Do not insert the calibration plate, as touching the circles may affect the accuracy of the calibration.
7.3 Repair / Troubleshooting
The optical 3D scanner contains sensitive components and sensor technology. Repair work may therefore only be carried out by CADstar
Technology service personnel.
In case of malfunctions that cannot be solved by restarting the scanner, contact customer service (see the chapter
"Customer service").
8. Disposal
The scanner can be returned to the dealer or manufacturer for disposal. Therefore, if you have any questions regarding proper disposal,
please contact our customer service department.
Please note that the scanner is an electronic device intended for use in commercial/industrial environments only.
Disposal via public waste management authorities is therefore not permitted.
Содержание CS.Neo2
Страница 1: ...www cadstar dental Instruction manual CS Neo2 CS Neo2 Pro CS Neo2 Max...
Страница 15: ...EN 14 9 1 Product diagrams clearances...
Страница 30: ...14 DE 9 1 Produktskizzen Abst nde...
Страница 45: ...IT 14 9 1 Schizzi del prodotto e distanze...
Страница 60: ...14 ES 9 1 Bocetos y espaciado de los productos...
Страница 75: ...FR 14 9 1 Sch mas de produits distances...
Страница 78: ...2 RU 1 3D CADstar Technology 1 1 1 1 1 CADstar 1 1 2 3D...
Страница 81: ...RU 5 1 8 1 9 1 10 1 11 3D 2...
Страница 82: ...6 RU 2 1 2 2...
Страница 83: ...RU 7 2 3 1 2 1 2 11 13 2 4 2cm 15cm 1 2 3D 3 4 5 6 7...
Страница 87: ...RU 11 5 1 24 5 2 5 3 1 6 USB 1 5 USB CADstar USB USB3 0 7...
Страница 88: ...12 RU 7 1 3D 7 2 7 3 3D CADstar Technology 8...
Страница 90: ...14 RU 9 1...
Страница 105: ...RO 14 9 1 Diagrame produs i aprob ri...
Страница 120: ...14 PL 9 1 Schemat produktu i dopuszczenia...
Страница 135: ...EN 14 9 1 r n diyagramlar ve a kl klar...
Страница 137: ...A 5500 Bischofshofen Tel 43 0 6462 32 880 office cadstar dental www cadstar dental...