19
ENGLISH
RECOMMENDED SEPARATION DISTAN-
CES BETWEEN PORTABLE AND MOBILE
RF COMMUNICATIONS EQUIPMENT AND
THE EQUIPMENT OR SYSTEM - FOR EQUI-
PMENT AND SYSTMES THAT ARE NOT
LIFE - SUPPORTING
For any question, please contact us, CA-MI Srl,
first, we will sort the problem for you, please refer
to the last page for contact information.
Recommended separation distances between portable and
mobile RF communications equipment and Infra-red Forehe-
ad Thermometer
Infra-red Forehead Thermometer is intended for use in an
electromagnetic environment in which radiated RF distur-
bances are controlled. The customer or the user of Infra-red
Forehead Thermometer can help prevent electromagnetic in-
terference by maintaining a minimum distance between por-
table and mobile RF communications equipment (transmit-
ters) and Infra-red Forehead Thermometer as recommended
below, according to the maximum output power of the
communications equipment.
Rated
maximum
output of
transmitter
W
Separation distance according to frequency
of transmitter m
150 kHz to
80 MHz
80 MHz to
800 MHz
800 MHz to 2.5
GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not
listed above the recommended separation distance d in
meters (m) can be estimated using the equation applicable
to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) accor-
ding to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for
the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.