Infra-red Forehead Thermometer is intended for use in the
electromagnetic environment specified below. The customer
or the user of Infra-red Forehead Thermometer should assure
that it is used in such an environment
.
Immunity
test
IEC 60601
test level
Com-
pliance
level
Electromagnetic
environment -
guidance
Electrostatic
discharge
(ESD)
IEC 61000-
4-2
±8kV con-
tact
±2kV, ±4kV,
±8kV
±15kV air
±8kV
contact
±2kV,
±4kV,
±8kV
±15kV air
Floors should be
wood, concrete
or ceramic tile. If
floors are cove-
red with synthe-
tic material, the
relative humidity
should be
at least 30%.
Electrostatic
transient /
burst
IEC 61000-
4-4
±2kV for
power sup-
ply lines
±1kV for
input/output
lines
Not
appli-
cable
Not
applicable
Surge
IEC 61000-
4-5
±1kV
differential
mode
±2kV com-
mon mode
Not
appli-
cable
Not
applicable
MANUFACTURER’S DECLARATION -
ELECTROMAGNETIC IMMUNITY
RF emissions
CISPR 11
Class B
Infra-red Forehead
Thermometer is
suitable for use in all
establishments, inclu-
ding domestic establi-
shments and those
directly connected to
the public low-voltage
power supply network
that supplies buildings
used for domestic
purposes.
Harmonic
emissions IEC
61000-3-2
Not applicable
Voltage fun-
ction / flicker
emissions IEC
61000-3-2
Not applicable