Radiated RF
IEC 61000-
4-3
3V/m
80MHz to
2.5GHz
3V/m
80MHz to
2.5GHz
80 MHz
to 800 MHz
385MHz-
5786MHz
Test speci-
fication for
ENCLO-
SURE POR
IMMUNI-
TY to
RF wireless
commu-
nication
equipmen-
t(refer to
table 9 IEC
60601-1-
2:2014)
385MHz -
5786MHz
Test speci-
fication for
ENCLOSURE
POR IM-
MUNITY to
RF wireless
communica-
tion equip-
ment(refer
to table 9
IEC 60601-
1-2:2014)
800 MHz
to 2.5 GHz
Where
p is the
maximum output
power rating of
the transmitter in
watts(W) according
to the transmitter
manufacturer and
d
is the recommended
separation distance
in metres(m).
b
Field strengths form
fixed RF transmitters,
as determined by
an electromagnetic
site survey,
a
should
be less than the com-
pliance level in each
frequency range.
b
Interference may oc-
cur in the vicinity of
equipment marked
with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic is affected by absorption and reflection
from structures, objects and people.
a
Field strengths from transmitters, such as base stations for
radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmit-
ters, an electromagnetic site survey should be considered. If
the measured filed strength in the location in which Infra-red
Forehead Thermometer is used exceeds the applicable RF
compliance level above. Infra-red Forehead Thermometer
should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be ne-
cessary, such as reorienting or relocating Infra-red Forehead
Thermometer.
b
Over the frequency range 150kHz to 80 MHz, filed strengths
should be less than 3V/m.