11.  Reverse steps to reinstall.
12.  Verify proper operation before returning the product to service.

Heater kit assembly

Tools required:

• 

#2 Phillips screwdriver

Procedure:

1.  Unplug the product.
2.  Drain the water out of the reservoir.
3.  Using a #2 Phillips screwdriver, remove and save the 9 screws that secure the tray to the reservoir. P/N 90018082 (x4)  

P/N 90018075 (x5). Save the screws and the tray assembly.

4.  Using a #2 Phillips screwdriver, remove the four screws from the back of the upper housing handle. Save the screws and 

upper housing handle.

5.  Remove the two wires from the Cartridge Heater. One goes to the PC Board, the other goes to a neutral wiring connector.
6.  Remove the Temperature Sensor (

Figure 5 on page 25

, item 8) from the Brass Heater Manifold.

7.  Cut the Gray Hose connected to the Brass Heater Manifold.
8. 

Remove the two brass fittings on the bottom of the Tray assembly which hold the brass manifold block to the tray.

9.  Discard the two O-rings between the brass heater manifold and the tray assembly.
10.  Remove and discard the heater kit assembly.
11.  Reverse steps to reinstall.
12.  Verify proper operation before returning the product to service.

Heater assembly

Tools required:

• 

#2 Phillips screwdriver

Procedure:

1.  Perform the steps for the Upper Front and Rear Housing.
2.  Remove the two wires from the Cartridge Heater. One goes to the PC Board, the other goes to a neutral wiring connector.
3. 

Remove the two brass fittings on the bottom of the Tray assembly that hold the brass manifold block to the tray.

4.  Pay special attention when reassembling the following:

a. 

Gray hose can be placed in warm water (<50 ºC) to facilitate installation on manifold fittings.

b.  All O-rings must be seated. The reservoir O-ring is shaped. The shape must be aligned with the reservoir.
c. 

All fittings on the manifold should have Teflon tape applied to prevent leaks.

d.  Make sure wires are routed away from the cooling fan blades.
e.  For proper screw size and location for reservoir and housing, refer to parts illustration and replacement parts list.
f.  Transfer the serial number to the new reservoir and cover the serial number label window

101176 Rev 1

Service

27

c2dx.com

Содержание T/PUMP TP700

Страница 1: ...www c2dx com 101176 Rev 1 2021 10 Rev 1 TP700 TP700C T PUMP SERIES LOCALIZED TEMPERATURE THERAPY SYSTEM Maintenance Manual...

Страница 2: ......

Страница 3: ...ngerous voltage Type BF applied part Medical Equipment Recognized by Underwriters Laboratories LLC With Respect to Electric Shock Fire and Mechanical Hazards only in accordance with UL 60601 1 2003 an...

Страница 4: ......

Страница 5: ...tional checks 14 Functional check and safety inspection form 15 Functional check setup Clik Tite connectors 16 Functional check setup Colder connection 16 Internal pump inspection 17 Flow and operatin...

Страница 6: ...eader of a potentially hazardous situation which if not avoided may result in minor or moderate injury to the user or patient or damage to the product or other property This includes special care nece...

Страница 7: ...to make sure the product operates as intended CAUTION Do not modify the product or any components of the product Modifying the product can cause unpredictable operation resulting in injury to patient...

Страница 8: ...aise or lower the temperature of local anatomical sites Indications for use The T Pump localized temperature therapy system is for use in supplying warm or cold water at controlled temperatures via th...

Страница 9: ...cryoglobulinemia and paraxysmal cold hemoglobinuria Application to body surface in patients with impaired sensation In combination with topical solutions whose toxicity may be affected by the applica...

Страница 10: ...0 F 48 C 20 F 28 C C2Dx reserves the right to change specifications without notice Product illustration A B C D E F G H I K J A Handle vents B Keypad C Flow indicator D Warm or cool delivery E Therapy...

Страница 11: ...it is on Yellow indicates power is supplied to the unit but the unit is not on F Therapy cycles 20 minute cycle 30 minute cycle or continuous cycle G Setpoint lock Prevents tampering of settings Conta...

Страница 12: ...8 Serial number location A Introduction 101176 Rev 1 c2dx com...

Страница 13: ...erial b Using a clean soft cloth moistened with a phenolic disinfectant or 10 bleach solution 3 Follow specified contact time in accordance with the manufacturer s instructions for use 4 Using a clean...

Страница 14: ...oir Assembly Includes Reservoir Gasket Rubber Feet Label 1 1 03394000 Fitting Manifold 2 2 03791001 Strap Assembly Hose 1 3 04152000 Fan 1 5 08086000 Hose Assembly Clik Tite Connectors 1 6 08648000 Ho...

Страница 15: ...y TP700 TP700C 1 38 100267000 Cord Power Assy TP700 TP700C 1 39 40 100269001 Membrane Panel Professional TP700 TP700C 1 41 100275000 Cap Tethered Rivet Assembly 1 42 101158000 Magnet Driver Assembly 1...

Страница 16: ...trouble free operation when the pad is folded The Mul T Pad series products with Clik Tite connectors can be interconnected to provide therapy to more than one body site at a time Mul T Pad Model Conn...

Страница 17: ...oven fabric on both sides 10 8002 062 622 For use with Colder connectors 22 in 56 cm 15 in 38 cm 30 Non woven fabric on both sides 10 8002 062 626 For use with Colder connectors 26 in 66 cm 18 in 46 c...

Страница 18: ...unctional checks Perform Functional Checks during maintenance or after you repair an internal component or as specified per your facilities preventive maintenance program This is to make sure you have...

Страница 19: ...tomer service Functional check and safety inspection form Inspection forms vary from hospital to hospital The following sample form is intended as a guide to record the important parameters T Pump Tem...

Страница 20: ...s on the supply side A B C D E Figure 1 Functional check setup for Clik Tite connectors 1 Plug into a properly grounded outlet A Figure 1 on page 16 2 Fitting and tube closest to the front of the T Pu...

Страница 21: ...90018075 x5 3 Perform a visual inspection of all internal parts 4 Remove any accumulated dirt with a vacuum cleaner or air compressor 5 Reverse steps to reinstall 6 Verify proper operation before retu...

Страница 22: ...rature reading every 1 2 minute for 5 minutes average the 10 readings and record the flow reading from the TPT9 Verify the average temperature is at 107 F 2 F 42 C 1 C and flow is 7 gph 26 5 lph minim...

Страница 23: ...will continue until the unit is put back into Standby mode 8 Press the On Standby button The T Pump Temperature Therapy Pump goes into Standby mode 9 Unplug the T Pump Temperature Therapy Pump 10 Unpl...

Страница 24: ...he pump performs a self test of the lights and audible alarm with a light pattern and a short audible beep The pump goes to Standby mode with only the Standby light on 7 Press the On Standby button 8...

Страница 25: ...is NOT spinning when the unit is plugged in and in run mode Otherwise check the impeller for any obstruction or replace the Impeller magnet assembly If the fan is spinning and there is no defect on t...

Страница 26: ...roximately 10 minutes Indication only to indicate an ON Therapy Cycle period is timing Indication only Both the Temperature and Therapy Cycle Setpoint lights are BLINKING Unit is in OFF Therapy Cycle...

Страница 27: ...Ice is DEPLETED Refill with room temperature water to proper level Reference Flow indicator light is ON section Drain excess water to Cooling water line and fill remainder of reservoir with ice Water...

Страница 28: ...To prevent unauthorized temperature setting changes a lockout key is available on the TP700 TP700C Professional model The limit thermostat Theory of operation on page 25 item 2 is on the brass manifo...

Страница 29: ...service items 1 PC board temperature controller 2 Manifold backup limit thermostat 3 Pump motor 4 Tray assembly 5 Cartridge heater 6 Brass manifold block 7 Return hose fitting 8 Temperature sensor 9...

Страница 30: ...e back of the upper housing handle Save the screws 5 Remove and discard the upper or rear front housing assembly 6 Reverse steps to reinstall 7 Verify proper operation before returning the product to...

Страница 31: ...manifold and the tray assembly 10 Remove and discard the heater kit assembly 11 Reverse steps to reinstall 12 Verify proper operation before returning the product to service Heater assembly Tools req...

Страница 32: ...PCB Kit B from the Front Housing C Figure 6 on page 28 This prevents interference between the ribbon and moving components A B C Figure 6 Ribbon cable 5 Verify proper operation before returning the pr...

Страница 33: ...101176 Rev 1 T Pump Temperature Therapy Pump assembly 29 c2dx com...

Страница 34: ...101176 Rev 1 T Pump Temperature Therapy Pump assembly 30 c2dx com...

Страница 35: ...sensor TP700 TP700C 1 3K 100819001 KIT Upper Rear Housing Assembly Includes Rear Plastic Housing with Label TP700 TP700C 1 4K 100818001 KIT Upper Front Housing Assembly Includes Front Plastic Housing...

Страница 36: ...14020 Screw type BT thread cutting 8 x 3 4 LG 4 22 90603000 Label ground identification 1 23 91275059 Screw machine flat C Sunk HD CR SST 6 32 x 3 8 LG 1 24 91390000 Toroid ferrite ring 1 25 91454052...

Страница 37: ...00378000 Cap for Impeller Pin 1 51 100578000 Label danger 1 52 100706000 Wiring harness 1 53 100706001 Wiring harness 1 54 100706002 Wiring harness 1 55 100144000 Heater assembly classic 120V 60Hz 1 5...

Страница 38: ......

Страница 39: ...C2Dx Inc 555 East Eliza Street Ste A Schoolcraft MI 49087 USA www c2dx com 101176 Rev 1 2021 10 Rev 1...