Problem
Possible Cause
Solution
T/Pump
®
Temperature Therapy
Pump
will not turn on
The electrical cord is NOT plugged
into a properly grounded Hospital
Grade receptacle.
Damaged cord or plug
Defective
PC Board / Sensor
Defective
Membrane Panel
Insert the
plug
into the properly
grounded Hospital Grade receptacle.
Visually check the
power cord
for any
defects. If defective, replace the
power
cord
.
Refer to
T/Pump
®
Temperature
Therapy Pump
will NOT heat.
Replace the
Membrane Panel.
T/Pump
®
Temperature Therapy
Pump
will not pump
Water level is LOW or reservoir is
EMPTY
Loss of POWER to motor
Defective Motor / Defective Impeller
magnet or magnet driver assembly
Refill with room temperature water to
proper level.
Check the POWER (TP700 / TP700C
- 120V ± 10% ) across the motor
connections. If present check for
defective motor / defective impeller
magnet or magnet driver assembly.
If there is NO power, replace the PC
Board.
Replace the motor if the fan is NOT
spinning when the unit is plugged in
and in
run mode
. Otherwise, check
the impeller for any obstruction
or replace the
Impeller
magnet
assembly.
If the fan is spinning and there is
no defect on the impeller magnet
assembly, check the
magnet driver
assembly
to see if the screw holding
the driver magnet to the motor shaft
is tight.
Warning indicator on with unit in
Standby Mode
Unit shut down in an OVER
temperature condition
Empty the reservoir and refill with
room temperature water.
Make sure all clamps are OPEN.
Press the
On/Standby
button.
Verify flow through the pad.
The
Warning
light will turn OFF within
5 minutes.
Temperature or Therapy Time buttons
do not work
The buttons have been LOCKED
Defective
Membrane Panel
Press and hold the lock button for wo
(2) seconds.
Replace the
Membrane Panel.
Flow indicator and Standby indicator
are ON with
T/Pump
®
Temperature
Therapy Pump
NOT pumping
Unit detected a Flow warning for
more than 5 minutes, thus goes to
Standby
Reference “Flow indicator light is ON”
Correct the problem, and press the
On/Standby
to put the unit back into
Run mode.
101176 Rev 1
Troubleshooting
21
c2dx.com
Содержание T/PUMP TP700
Страница 2: ......
Страница 4: ......
Страница 12: ...8 Serial number location A Introduction 101176 Rev 1 c2dx com...
Страница 33: ...101176 Rev 1 T Pump Temperature Therapy Pump assembly 29 c2dx com...
Страница 34: ...101176 Rev 1 T Pump Temperature Therapy Pump assembly 30 c2dx com...
Страница 38: ......
Страница 39: ...C2Dx Inc 555 East Eliza Street Ste A Schoolcraft MI 49087 USA www c2dx com 101176 Rev 1 2021 10 Rev 1...