Problem
Possible Cause
Solution
defective. Replace
thermostat
if it
does not pass Backup Limit
Thermostat Test from Functional
check.
If there is NO Power to the heater,
replace the
PC Board.
T/Pump
®
Temperature Therapy Pump
will NOT Cool
Reservoir is EMPTY
Flow is BLOCKED
Ice is DEPLETED
Refill with room temperature water to
proper level
Reference “Flow indicator light is ON”
section
Drain excess water to
Cooling
water
line and fill remainder of reservoir with
ice.
Water LEAKS from hose connectors
Replace
Clik-Tite
®
connector
Male: P/N: 03887001
Female P/N: 03884001
Snap
Clik-Tite
®
connector
SHUT
Secure pad connection to pump
(See quick disconnects). Replace
connectors or pad if defective.
Wiring diagram
101176 Rev 1
Troubleshooting
23