Problem
Possible Cause
Solution
Flow indicator light is ON
Water flow to pad or hose is
RESTRICTED
Clamp is CLOSED
Water level is low or reservoir is EMPTY
T/Pump
®
Temperature Therapy Pump
is filled with water that is too HOT
Straighten the hose
Open the clamp
Refill with room temperature water to
proper level
Refill with room temperature water to
proper level
Warning indicator & Audible alarm
(Flash / Beep)
A High Heat (107°F / 42°C) or Cooling
Setpoint was selected (50°F / 10°C)
Loss of Power while unit was in a
Therapy mode.
(Possible Power Fail)
Unit is running after a 20- or 30-
minute “OFF” Therapy Cycle period,
has reached the desired Setpoint, and
is now timing the 20- or 30- minute
ON cycle period
The unit just went into, or came out of
Lock mode
Indication only: A Setpoint outside
body temperature range is selected.
Insert the plug fully into the receptacle,
place the unit into
Standby mode
, then
unplug the
T/Pump
®
Temperature Therapy Pump.
If Power is removed while unit is in On-
Mode, the
Power Fail alarm
will beep
for approximately 10 minutes.
Indication only to indicate an “ON”
Therapy Cycle period is timing.
Indication only.
Both the Temperature and Therapy
Cycle Setpoint lights are BLINKING
Unit is in “OFF” Therapy Cycle time
Indication only.
Temperature Setpoint light BLINKING Unit is warming up to the selected
Setpoint
Unit is in
Cooling mode
, for longer
than 40 minutes
Indication only.
Follow the “shutdown” procedure.
Drain the water in reservoir to ice fill
level, and refill with ice. Follow the
“start-up” procedure.
T/Pump
®
Temperature Therapy
Pump
will NOT Heat
Reservoir is EMPTY
Flow is BLOCKED
Therapy Cycle is OFF
Heater is NOT receiving Power, or is
defective
Refill with room temperature water to
proper level
Reference “Flow indicator light is ON”
above
Wait for Therapy Cycle ON
Check if the
Set Point
is set too low.
Remove
cover
and check Power
(TP700 / TP700C - 120V ± 10% ) to the
heater
at J2 on the PCB.
NOTE
: Power
to J2 cycles ON and OFF
If Power is present and thermostat is
CLOSED, heater is probably defective.
Cold heater resistance is approximately
49 ohms (120V). Replace
heater
if
required. Perform
Functional checks on
page 14.
If Power is present and thermostat is
OPEN, thermostat is probably
101176 Rev 1
Troubleshooting
22
c2dx.com
Содержание T/PUMP TP700
Страница 2: ......
Страница 4: ......
Страница 12: ...8 Serial number location A Introduction 101176 Rev 1 c2dx com...
Страница 33: ...101176 Rev 1 T Pump Temperature Therapy Pump assembly 29 c2dx com...
Страница 34: ...101176 Rev 1 T Pump Temperature Therapy Pump assembly 30 c2dx com...
Страница 38: ......
Страница 39: ...C2Dx Inc 555 East Eliza Street Ste A Schoolcraft MI 49087 USA www c2dx com 101176 Rev 1 2021 10 Rev 1...