Remove product from service before you perform the preventive maintenance inspection. Check all items listed during
annual preventive maintenance for all C2Dx
®
products. You may need to perform preventive maintenance
checks more often based on your level of product usage. Service only by qualified personnel.
WARNING
Risk of electric shock. Disconnect power before servicing or cleaning the
T/Pump
®
Temperature Therapy Pump.
Note:
Change the water monthly or more often depending on use.
Inspect the following items:
Line cord, the plug, and the housing for cuts or cracked insulation and quality of the strain relief, if damaged
replace
Membrane panel for cuts or cracks, if damaged replace
Plastic parts on the pump for cracks
Connectors for cracks and missing O-rings, replace connectors if necessary
Hose connections and connectors function properly
Pads and hoses are free from tears, cracks, and water leaks
Perform
Functional checks on page 14.
Inspection
Value
Okay
Action Needed
Action Taken
Measure current
leakage <100 μA
Product serial number:
Completed by:
Date:
Functional checks
Perform Functional Checks during maintenance, or after you repair an internal component, or as specified per
your facilities preventive maintenance program. This is to make sure you have optimum performance, dependability, and
safety of your product.
WARNING
•
Only qualified service personnel should perform this Functional Check. Test procedures completed improperly may result
in equipment damage.
•
Do not perform functional checks with an empty reservoir as this may result in damage to the product.
•
Always perform a functional check after any field service to make sure the product operates as intended.
This section provides a complete check of all
T/Pump
®
Temperature Therapy Pump
functions. See
(Functional check and safety
inspection form on page 15)
to record test data. Follow the steps in this section. Pay attention to each step, and the expected result. If at
any time you cannot get the expected result, press the
On/Standby
button to stop the test, then restart the procedure. If after a second
attempt you cannot get the expected result, press
On/Standby
button to stop the test. Unplug the
T/Pump
®
Temperature Therapy Pump
and call your dealer or contact C2Dx
®
Technical Support Department for help.
101176 Rev 1
Preventive maintenance
14
c2dx.com
Содержание T/PUMP TP700
Страница 2: ......
Страница 4: ......
Страница 12: ...8 Serial number location A Introduction 101176 Rev 1 c2dx com...
Страница 33: ...101176 Rev 1 T Pump Temperature Therapy Pump assembly 29 c2dx com...
Страница 34: ...101176 Rev 1 T Pump Temperature Therapy Pump assembly 30 c2dx com...
Страница 38: ......
Страница 39: ...C2Dx Inc 555 East Eliza Street Ste A Schoolcraft MI 49087 USA www c2dx com 101176 Rev 1 2021 10 Rev 1...