- 23 -
Appendix Description of status code and solution
Status code
Description
Solution
1
The position of the displacement sensor was
incorrect when the test was started.
Adjust the position of the displacement sensor,
make the amount of deformation is displayed
between 1 and 2 mm.
2
The order of the feedback voltage value of
the pressure correction table is wrong.
Arrange the feedback voltage values of the
pressure correction table in order from small to
large.
3
The actual pressure value sequence of the
pressure correction table is wrong.
Arrange the actual pressure values of the
pressure correction gauge in order from small to
large.
4
The voltage value sequence of the
displacement calibration table is wrong.
Arrange the voltage values of the displacement
calibration table in order from small to large.
5
The displacement value sequence of the
displacement calibration table is wrong.
Arrange the displacement values of the
displacement calibration table in order from
small to large.
6
During pressure correction, the pressure
sensor is overloaded.
Reversely turn the electronic handwheel to lower
the lifter, and then it can automatically recover.