Do not operate this device in the presence of nitrous oxide and oxygen.
Keep away from toxic or hazardous steam.
Do not use this device if the room temperature is higher than 35
℃
(
95
°F)
. If the
ambient temperature of the room is higher than 35
℃
(
95
°F),
the air flow from the
device may rise to exceed 40
℃
(105
°F),
which may cause damage to the user’s
airway.
Do not use this device in direct sunlight or near heating equipment. Doing so
may increase the air temperature of the output to unsafe levels.
To avoid electrical shock, disconnect the power cord before cleaning. Do not
immerse the device in fluid of any kind.
Contact your healthcare provider if symptoms of sleep apnea recur.
Regularly check the power cable and accessories for any damage or wear.
Disconnect the power supply before checking the device.
When using this device, make sure the mask is positioned above the height of
the host unit. Failure to do so may cause any condensed water in the tubing to flow
into the user’s nose, creating the risk of possible suffocation.
Stop using this device if the humidifier is damaged.
Do not touch the heater plate after turning the device off. Let the heater plate
cool off before touching. The plate will remain hot for several minutes after the
power supply has been disconnected.
Do not add water hotter than
35
℃
(
95
°F)
.
Do not splash any water into the device when refilling the water tank.
Be sure that the device is in a place that cannot be accessed or touched by
children, who could be entangled, injured, or strangled by the tube.
Appliance coupler is used as isolation means for the device, please do not
position the device so that it is difficult to operate the coupler.
Do not attempt to modify this device without written authorization from the
manufacture.
Do not place the device next to any curtains or similar obstruction. Curtains can
block the airflow to the device, which may cause this device to overheat.
Do not block the air Intake Port. Doing so will interfere with the therapy.
The proper placement and positioning of the mask on the face is critical to the
consistent operation of this device.
The local distributor of this device is available to provide technical support.
Do not perform any maintenance on the device while user is receiving the
treatment.
All materials that come in contact with the human body have been tested for
biocompatibility and meet the requirements of biological compatibility
Approved adapters are specified as part of the device.
1.3 Cautions
The following precautionary statements indicate conditions or actions that might
damage the device or injure the user. Please read this section carefully.
Do not begin wearing the mask until the device has started to run normally.
Do not operate the device if the ambient temperature is outside of the operating
temperature range (listed under Section 13). If the device has been exposed to
temperatures that are above the operating temperature range, return the device to
room temperature before use.