![Byond ResPlus C-20A CPAP Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/byond/resplus-c-20a-cpap/resplus-c-20a-cpap_user-manual_2837202025.webp)
22
The device is
excessively loud.
The tube isn’t connected
properly. The mask or tubing
may also be leaking.
Reconnect the tubing and
ensure there are no leaks in
the tubing.
The pressure can’t
be set.
The Ramp feature is on.
Turn off the Ramp feature,
then re-set the pressure.
If the problem can’t be eliminated immediately by the above suggestions, please
contact your device supplier for repair.
To avoid more injuries or faults, please don’t use the device when any of these
problems persist.
Opening this device voids the warranty. Only professionals certified by BEYOND
Medical are authorized to open the device.
13. Specifications
Environmental
Operating
Storage
Temperature
5℃ to 35℃ (41˚F to 95˚F)
-20℃ to 55℃(-13˚F to 158˚F)
Humidity
15% to 93% (no condensation)
15% to 93% (no condensation)
Atmospheric
Pressure
700 to 1060 hPa
700 to 1060 hPa
Physical
Dimensions
165mm*200mm*120mm,
280mm*200mm*140mm(with humidifier)
Weight
1.2Kg,2.2Kg (with humidifier)
Electrical
Power adapter
Input: AC100-240V, 50/60Hz,
1.8Amax
Output: DC24V, 3.75A
Type of protection Against Electric
Shock
Class II Equipment
Degree of protection Against Electric
Shock
Type BF applied part
Degree of protection Against Ingress of
Water
IP22
Mode of operation
Continuous
Applied part
SpO2 Probe, Tubing, Mask
Noise level