13
*Press
4-20
cmH2O
This setting sets the output pressure for the device. It
can be set in increments of ±0.5 cmH2O.
*StartPress
4.0-20.0
cmH
2
O
CPAP Mode
: The initial air pressure for the Ramp
Function is equal to or less than the standard air
pressure selected by the user.
APAP Mode
:
1.
When the initial air pressure is equal to or
less than the minimum pressure, it is used as
the starting pressure to begin the Ramp
Function.
2.
When the initial pressure is greater than the
minimum pressure, it is used as the output
pressure after power is turned ON.
This parameter can be set in increments of ±0.5
cmH
2
O.
*MinPress
4.0-20.0
cmH
2
O
The minimum output pressure parameter can be
auto-adjusted at increments of ±0.5 cmH
2
O.
*MaxPress
4.0-20.0
cmH
2
O
The maximum output pressure parameter can be
auto-adjusted at increments of ±0.5 cmH
2
O.
*Belex
0-3 level
Setting the Belex function while the device is in CPAP
Mode will enable the device to detect the respiratory
rhythm of the user. This function will lower the
pressure in the mask during exhalations to make the
user feel more comfortable.
Ramp
0-60 min
This parameter sets the time it takes for the air pressure
to ramp up to the final pressure selected by the user
(MaxPress)
Humidifier
0 (OFF)
to 5
(Maximu
m)
This parameter sets the output air humidity level for
the device. The higher the value, the higher the
humidity.
"0" indicates that the humidifier is off.
The default value before dispatched from the
manufacturer is “0”.
Tips
"*" indicates that this parameter can only be adjusted while the
parameter interface is unlocked. To unlock the parameter interface,
position the cursor over the “Parameter” icon and then press and
hold the Dial Button for 5 seconds to unlock.
9.4 System Setting Interface
Rotate the "Dial" button so that the cursor on the Main Interface screen is pointing at
“System”, then press the "Dial" button to enter into the System Setup Interface screen,
as shown in
Fig.9.4a .