10. Prompt
Prompt
Message
Description
Power
Failure
If the power supply is disconnected while the device in the
Working State , the device will provide a voice prompt via the
buzzer for approximately 30 seconds. Pressing the "On/Off "
Button or re-powering the device will stop the voice prompts.
After power is returned to the device, it will return to the normal
state.
LOW
VOLTAGE!!
When the device detects that the adapter output voltage is below
acceptable limits, the device will provide a voice prompt via the
buzzer and "LOW VOLTAGE!!" will appear on the display. The
warning sounds can be cancelled by pressing the Mute Button.
AIRWAY
BLOCK!!
When the air flow of the device is blocked, the device will
provide a sound prompt via the buzzer, and "AIRWAY
BLOCK!!" will appear on the display. The warning sound can be
cancelled by pressing the Mute Button.
HIGH
LEAKAGE!!
When there is a leak in the device's breathing tube, the device will
provide a sound prompt via the buzzer and "HIGH LEAKAGE!!"
will appear on the display. The warning sound can be cancelled
by the Mute Button.
HUMI
FAILURE!
When the device is connected to the humidifier, the device is
powered on, and the humidifier has been heating for 2 minutes,
but the humidifier is not working, the device will provide a sound
prompt via the buzzer and "HUMI FAILURE!" will appear on the
display. The warning sound can be cancelled by pressing the
Mute Button.
TF CARD
FULL!
When the TF card that has been inserted into the device reaches
its capacity limit, the device will provide a sound prompt via the
buzzer and "TF CARD FULL!" will appear on the display. The
audio prompt for this warning is a single tweet, it will only tweet
once.
11. Cleaning, Disinfection and Maintenances
11.1 Timing between Cleanings
Based on hygienic reasons, under normal usage, it is recommended that the operator
or user service and / or replace the parts as shown below:
•
Clean the device, tube and filter before the device is used for the first time.
•
Empty the water tank and clean it daily to prevent bacterial growth.
•
Clean the air filter at least once every 2 weeks.
•
Replace the air filter with a new filter at least every 3 months.
•
Replace the mask with a new one every 3 months.