29
GasAlertQuattro
Device Configuration
Stealth
When the
Stealth
option is enabled, the following features are disabled:
• backlight,
• audible alarms,
• visual alarms,
• IntelliFlash, and
• confidence/compliance beep
Only the vibrator and the LCD readings activate during an alarm condi-
tion.
When
Stealth
is enabled,
displays every 3 seconds between each
sequence of three heartbeats.
If an alarm occurs,
ceases to display until the gas concentrations
are below the low alarm setpoint.
The detector is shipped with the
Stealth
option disabled.
Datalog Interval
The
Datalog Interval (seconds)
field defines how often the detector
records a datalog (every
1
-
120
seconds). Enter the desired value.
The total number of 8-hour days datalogs that can be recorded is assum-
ing 90% of the day has no gas concentrations.
When the memory is full, the detector replaces the oldest datalogs with
the most recent datalogs.
The detector is shipped with the default setting of
5
seconds.
Datalog Interval
Total Number of Days
Datalogs Can Be Recorded
5 seconds
up to 15 days
15 seconds
up to 45 days
60 seconds
up to 180 days