WA-1150 and WA-1650 Wavemeter Operating Manual
3 Remote Communications
3-12
(09042-M-00 Rev C) EXFO Burleigh Products Group Inc Inc.
:MINimum[:STATe] {ON | OFF}
Description
Sets the state of the :MINimum function to either ON or OFF.
*RST value
OFF
Examples
:CALC:MIN ON
:CALC:MIN:STAT OFF
:RESet
Description
Resets the minimum, maximum and start values to the current reading.
Resets the deltas to zero and sets the elapsed time to 00:00:00. This affects
the CALCulate subsystem as well as the instrument's front panel display. It
is the same as using the Reset function through the display.
Examples
:CALC:RES
:STARt[:STATe]?
Description
Queries the state of the :STARt function.
Query Response
{0 | 1} Where 0 indicates OFF and 1 indicates ON.
Examples
:CALC:STAR?
:CALC:STAR:STAT?
:STARt[:STATe] {ON | OFF}
Description
Sets the state of the :STARt function to either ON or OFF.
*RST value
OFF
Examples
:CALC:STAR ON
:CALC:STAR:STAT OFF
:TIME[:ELAPsed]?
Description
Queries the elapsed time since the instrument was turned on or was reset.
Query Response
A time in HH:MM:SS format.
Examples
:CALC:TIME?
:CALC:TIME:ELAP?