WA-1150 and WA-1650 Wavemeter Operating Manual
3 Remote Communications
EXFO Burleigh Products Group Inc. (09042-M-00 Rev C)
3-19
11
2048
Not used.
12
4096
Input laser fringe strength is too high.
13
8192
Reference laser % of peak power is out of range
(WA-1650 only).
SYSTem Subsystem
The SYSTem subsystem supplies information and status about the SCPI command parser
and allows control over the instrument's front panel.
SYSTem Subsystem Functions
:ERRor?
Description
Reads error strings from the SCPI Error Queue. If the Error Queue has any
entries, the Error Queue bit is set in the Status Byte. The instrument has a 30
entry, first-in, first-out queue. Repeatedly sending the query :SYST:ERR?
returns the error numbers and descriptions in the order in which they
occurred until the queue is empty. Any further queries return 0, "No error"
until another error occurs. A table of possible errors is shown below.
Query Response
<integer>, <string> (ex. –104, “Data type error”)
Examples
:SYST:ERR?
SCPI Errors
Error Number Description
0
No error
-101
Invalid character
-102
Syntax error
-103
Invalid separator
-104
Data type error
-220
Parameter error
-221
Settings conflict
-222
Data out of range
-230
Data corrupt or stale