WA-1150 and WA-1650 Wavemeter Operating Manual
3 Remote Communications
3-6
(09042-M-00 Rev C) EXFO Burleigh Products Group Inc Inc.
summary bit will be set in the STB.
Query Response
Returns an integer which is the sum of all the bit values for those bits that are
set (see Event Status Register table below).
Event Status Register
Bit
Bit Value
Condition
7
128
Power On (PON)
6
64
Not Used
5
32
Command Error (CME)
4
16
Execution Error (EXE)
3
8
Device Dependent Error (DDE)
2
4
Query Error (QYE)
1
2
Not Used
0
1
Operation Complete (OPC)
*IDN?
Description
The *IDN (identification number) query returns a string value which contains
the instrument type, serial number and firmware version. The third value is
the instrument serial number. The last value is the software version and will
reflect the actual version number.
Query Response
BURLEIGH WAVEMETERS, WA-1150, 1001, 1.0
*RCL
Description
The *RCL (recall) command restores instrument settings. It is the same as
using the Restore All function through the display.
*RST
Description
The *RST (reset) command returns the instrument’s settings to a known state.
Commands affected by *RST
Command
Setting after an *RST command
:CALCulate:DELTa[STATe]
OFF
:CALCulate:DELTa:METHod
STARt
:CALCulate:MAXimum[:STATe] OFF
:CALCulate:MINimum
OFF
:CALCulate:STARt[:STATe]
OFF
:DISPlay:DIALog
MAIN