SCS 104
lower alarm limit (alarm low)
Display → Cooler →
A Lo
Here you can set the lower threshold for the visual signal and the alarm relay. The alarm limit is set
based on the cooler temperature setting.
Parameter range:
-1 °C to -3 °C (-1.8 °F to -5.4 °F)
Factory setting:
-3 °C (-5.4 °F)
Note:
This menu item is hidden if the keylock is enabled.
Exit submenu 1
Display → Submenu →
E
Selecting this will return you to the main menu.
5.2.4.3 Submenu global settings
Temperature unit
Display →
toP
→
unit
Used to select the temperature display unit.
Parameter range:
°C
,
°F
Factory setting:
°C
Calibrate moisture detector
Display →
toP
→
h2o
(h2o)
If a moisture detector is installed, calibration can now be performed. To do so, the unit must be
flushed with dry gas.
Note:
Calibration was performed at the factory using ambient air. After replacing the moisture detector a
calibration is again required.
Calibrating the moisture detector will set the menu
sEn
to
hi
.
This menu will be hidden if the menu is locked.
If the unit has multiple moisture detectors built in, they will be numbered in the menu. In this case,
h2o
indicates the first,
h2o2
the second moisture detector. The same applies to setting the sensor sensitivity in menu
SEn
.
Moisture detector sensitivity
Display →
toP
→
SEn
If moisture detectors are installed, the sensitivity can be reduced here.
Parameter range:
hi
: high sensitivity
Lo
: low sensitivity
non
: no moisture detector
Factory setting:
hi
Note:
This menu will be hidden if the menu is locked.
14
Bühler Technologies GmbH
BECS0031 ◦ 02/2020