Operation
The product is operated via the SX-Suite software on a computer. See software oper-
ating instructions.
6.1
Journal button
Pressing the button generates a journal entry.
6.2
Entering reference data in the journal
To perform a calibration and continuously check the calibration, reference data is re-
quired.
Continuous checking of the calibration is performed according to the requirements of
the production process.
NOTE
The journal entry is identified by date and time.
u
Press and hold the button for one second.
ð
The connected software creates a journal entry.
u
Remove the sample at the sample removal point.
u
Mark sample with date, time and sensor number.
u
Carry out a laboratory analysis.
u
Insert the reference data in the journal for creating the calibration model. See
"SX-
Suite User Manual "
and
"SX-Plus User Manual "