Fig. 7: Fixing sensor with screws
Tools required:
Torque wrench, size Torx T30
Tightening torque: 8.4 Nm +-1
Precondition:
R
Installation point has been established. See Chapter 5.5 "Establishing installation
point", page 21.
R
The fixing points or bolts conform to M6 A2-70 15 mm
u
Fix the sensor to the installation point using the bolts.
5.7
Connecting the sensor
Tools required:
Torque wrench, size 7 mm AF
Torque wrench, size Torx T20
NOTE
Make sure that the power is not switched on when connecting the sensor.
NOTE
Loss of performance due to use of unsuitable device cables
Max. cable length between installation box and sensor 10 m.