NIR-Online GmbH
Product description | 3
Operation Manual Process Analyser PA2
15/36
3.6
Technical data
3.6.1
Sensor
Specifications
PA2
Dimensions (W x D x H)
236 x 278 x 252 mm
Weight
14 kg
Max. working pressure
30 bar at flange
Coolant inlet temperature
+10 °C to +30 °C
Coolant flow rate
10 L/h
Coolant pressure
max. 0.5 bar
Product temperature
(temperature at flange with water cooling)
-10 °C to +70 °C
Product temperature
(temperature at flange without water cooling)
-10 °C to + 40 °C
Vibrations
0.2 G at 0.1 - 150 Hz
Wavelength spectrum
400 - 2200 nm
Detector
Diode array
Average measurement time
5 spectra/s
Protection class
IP66
Lamp type
Tungsten-halogen lamp
Number of lamps
2
Service life of lamp
9000 h
3.6.2
Installation box
Specifications
Installation box
Dimensions (W x D x H)
300 x 300 x 167 mm
Weight
(excluding cables)
6 kg
Weight
(inc. cables, 2 x 10 m)
7.4 kg
Frequency
50/60 Hz
Power consumption
30 W
Power supply
110 - 240 ± 10 % VAC
3.6.3
Ambient conditions
Max. altitude above sea level
2500 m
Ambient temperature
-10 °C ≤ Tamb ≤ + 40
°C
Max. relative air humidity
< 90 % non-condensing
Storage temperature
max. 45 °C