37
BÜCHI
Fat Determination
B-820/B-821
Gas flow setup
User configuration
Printer configuration
System configuration
Autosampler
Main menu
Autosampler:
yes
Injections:
2
Main menu
Preheating
p H2
55kPa
Sample weight/Mode
p air
50kPa
Determination
T oven
130°C
Data manager
T inj.
220°C
Setup
T FID
260°C
5 Operation
Baudrate:
9600
Data Bits:
8
Stop Bits:
1
Parity:
None
Protocol:
None
Main menu
PC-Interface
This menu is meant for the setup of interface parameters for
the PC interface. The cursor is positioned on
Baudrate
.The
Baudrate cannot be changed.
Position cursor on
Data bits
. Press the OPTION key to select
the number of bits for the transmission.
Position cursor on
Stop bits
. Press the OPTION key to select
<1> or <2>. Press the OPTION key to position
parity
even/
odd or none.
Position
protocol
on Software Handshake none or Hard-
ware Handshake RTS/CTS or none.
Position cursor on <Main menu>.
5.2.4 Autosampler configuration
Position cursor on <Setup>.
The cursor is positioned on <Gas flow setup>.
Position cursor on <Autosampler>.
The cursor is positioned on <
Autosampler
>. Press OPTION
key to select <
Manual injection
>
(no)
or <
Autosampler
>
(yes)
.
Position cursor on <
Injections
>. The number of injections per
sample weight may be selected from 1 to 9.
«Total fat» option: max. 9 injections per sample weight.
«Fatty acids» option: max. 6 injections per sample weight.
Position cursor on <Main menu>.