BTL BTL-07p Скачать руководство пользователя страница 26

BTL – 07p  

operating manual & user's guide 

page 26 

List of program numbers for combined ultrasound therapy/electrotherapy 

51. Lumbalgie 

Emitter head: 4 cm

2

. Frequency: 3 MHz, 1:4 PPR. Intensity: 0.5 W/cm

2

Electrotherapy: TENS cont., f = 100 Hz constant (program 1625). 
Contralateral electrode placement. 
Intensity above threshold sensitive. 
Application time: 1 - 5 minutes, step 1 minute, daily. Total number of treatments: 3 - 5. 

52.  Myalgia mm. superficiales 

Emitter head: 1 cm

2

. Frequency = 3 MHz, 1:2, 1:4 PPR. Intensity: 0.5 W/cm

2

Electrotherapy: DF, base 1 mA (program 0113).  
Contralateral electrode placement. 
Intensity above threshold sensitive. 
Application time: 1 - 2 minutes, max. 6 minutes, daily. Total number of treatments: 2 - 3. 

53.  Myalgia mm. profundi 

Emitter head: 1 cm

2

. Frequency = 1 MHz, 1:2, 1:4 PPR. Intensity: 1.5 - 1.7 W/cm

2

, step 0.1 W/cm

2

Electrotherapy: DF, base 1 mA (program 0113). 
Contralateral electrode placement. 
Intensity above threshold sensitive. 
Application time: 1 - 2 minutes, max. 6 minutes, daily. Total number of treatments: 3 - 5. 

54.  Relaxatio mm. superficiales 

Emitter head: 1 cm

2

. Frequency = 3 MHz, 1:1, 1:2, 1:4 PPR. Intensity: 0.5 - 1.0 W/cm

2

, step 0.1 W/cm

2

Electrotherapy: TENS cont., f = 100Hz (program 1725)  
Contralateral electrode placement. 
Intensity above threshold sensitive. 
Application time: 1 - 10 minutes, step 1 minute, daily. Total number of treatments: 1 - 6. 

55.  Relaxatio mm. profundi 

Emitter head: 4 cm

2

. Frequency = 1 MHz, continual PPR. Intensity: 0.5 - 1.0 W/cm

2

, step 0.1 W/cm

2

Electrotherapy: AMP, FM 100 - 200 Hz, steps 3/3 (program 0947)  
Contralateral electrode placement. 
Intensity at threshold motor. 
Application time: 1 - 10 minutes, step 1 minute, daily. Total number of treatments: 1 - 6. 

56.  Trigger points superficialis 

Emitter head: 1 cm

2

. Frequency = 3 MHz, 1:2 PPR. Intensity: 0.5 W/cm

2

Electrotherapy: TENS cont., f = 100 Hz const. (program 1625)  
Contralateral electrode placement. 
Intensity at threshold motor. 
Application time: 1 minute, daily. Total number of treatments: 1 - 5. 

57.  Trigger points profundi 

A/ Trigger point finding 

Emitter head: 1 cm

2

 (a 4 cm

2

 head can also be used, but proper aim is difficult and degree of accuracy is 

lowered). Frequency = 1 MHz, 1:4 PPR. Intensity: 2.0 W/cm

2

Electrotherapy: TENS continuous, imp. 0.05 ms, asymmetrically biphase, f = 100 Hz, (program 1525) 
Procedure: 
First apply contact gel (electrically conductive) to the examined area. Set the ultrasound parameters. 
 
Outside the hyper-allergic zone (HAZ) or the assumed trigger (TP) set the intensity of the electrical to threshold 
sensitive and through a fluent head movement we move to the assumed and reflective changed area. HAZ is 
announced by the patient as a increased intensity of a feeling by electric current bypass. TP localisation can be 
found in two ways: 

1. 

 

The patient reports a typical radiating pain. 

2. 

 

Asymptotic motor effect (muscle contraction) in the affected muscle. 

 

Содержание BTL-07p

Страница 1: ...v2 10z6BJ24 03 2004EN operating manual user s guide ...

Страница 2: ...nty 8 Maintenance and transport 8 Troubleshooting 8 Technical specifications of probe 9 Technical specifications of unit 9 Pulse width 10 II USER S GUIDE 11 Mechanism of action 11 Methods of application 12 Ultrasound frequency 13 Modes of operation 13 Emitter head size 13 Application time 13 Intensity 13 Procedure frequency 13 Total number of treatments 13 Specific contraindications of ultrasonic ...

Страница 3: ...18 Mechanism of action 18 Application time 18 Intensity 18 Total number of treatments 18 Frequency of treatment 18 Procedure prescription 18 A therapeutic example 18 III LIST OF PROGRAM NUMBERS FOR ULTRASOUND THERAPY 19 IV LIST OF PROGRAM NUMBERS FOR COMBINED ULTRASOUND THERAPY ELECTROTHERAPY 25 ...

Страница 4: ...by two pilot lights 7 Push button to set acoustic output 8 Push button to change indication of acoustic output 9 Duty Cycle push button 10 Column lights indicating contact of emitter head with skin 11 Small ultrasound emitter head connector 12 Signal lights Ready Emitter heads connected intensity therapy time is set Error Light signal a hardware defect indicating Power Light signal indicating the ...

Страница 5: ... connecting and identifying the emitter head The red diode on the emitter head indicates insufficient contact between the emitter head and the tissue If contact is insufficient the control light lights up red The countdown on the TIME display 3 stops Power absorption indicated by the light blinking stops Therapy is interrupted until contact is restored If there is proper contact the control signal...

Страница 6: ...f intensity is an area which is exposed to intensity higher than the preset intensity level There is an increased local temperature in these areas BTL 07p emitter heads have a very low BNR Further more the BTL 07p features an optic indicator to signal stationary waves placed on the Bar graph 10 the orange LED diode Respond to this signal by immediately removing the emitter head from the applicatio...

Страница 7: ...tly follow the safety instructions below 1 Read the operation manual carefully before you use the instrument for the first time 2 Make sure that the parameters of the mains correspond to those of the instrument 3 The mains must be installed and correspond to valid norms Call a qualified electrician if in doubt 4 If the instrument is exposed to a sudden temperature charge allow at least 1 hour for ...

Страница 8: ...he Manufacturer shall not be liable for incidental consequential or special damages arising out of or in connection with product use or performance except as may be otherwise accorded by law This warranty may differ from the warranty terms and conditions provided by your supplier and by applicable laws in your country Maintenance and transport The unit can be cleaned with a damp cloth Do not use a...

Страница 9: ...ning up and down After certain time you will hear a sound Release the Frequency button and wait until the bar graph LEDs will stop running the calibration is done You might control it after begin generating and leave the emitter head in the air the red LED should be on put the emitter head into the water the green LED should be on immediately READY signal does not light up Set application time and...

Страница 10: ...trols ON OFF switch marking 0 1 Standard accessories 1 ultrasound probe with area 1 cm 2 1 ultrasound probe with area 4 cm 2 1 bottle ultrasound gel 2 spare fuses external Pulse width ti ms tp ms T ms f Hz Duty factor 10 10 20 50 1 2 5 15 20 50 1 4 2 5 17 5 20 50 1 8 1 25 18 75 20 50 1 16 5 5 10 100 1 2 2 5 7 5 10 100 1 4 1 25 8 75 10 100 1 8 0 625 9 375 10 100 1 16 Important Each BTL 07p instrume...

Страница 11: ...nic head when the frequency is fixed A good quality head should not exceed 6 With a preset intensity of 1 W cm2 the intensity should not exceed 6 W cm 2 in any part of the ultrasonic beam The BNR of older ultrasonic heads as well as many of the newer ones where the producer does not mention the BNR is often 20 or more The Effective Radiating Area ERA is always smaller than the actual surface area ...

Страница 12: ...er head The therapist moves the emitter head in a circular pattern The application time is prolonged proportionally to the ERA and the size of the application site B Application site 1 Local application The ultrasound is applied to the affected site This is the most common type of application and is particularly suitable for the management of muscle spasm chronic post traumatic edemas etc 2 Segmen...

Страница 13: ...o large areas is difficult for the therapist lasts too long and the dose is not applied consistently Small ERA 1 cm2 emitter heads are used either to manage trigger points and scars or to manage small uneven surfaces which have to be treated by the subaqual method if the emitter head size is large Application time Individual application times vary They largely depend on the stage of the disease or...

Страница 14: ...forehead lobe as the calcified atheroma plate was disturbed It is widely believed that this was not an accident This information is included only as a strong reminder to the therapist that ultrasound is not to be used in these areas A patient must never apply ultrasound to himself or herself even if the application site can be reached easily Peripheral Nerves situated just below the surface on the...

Страница 15: ...nless such information is specifically requested However be sure to ask if epistaxis is present or if patient is menstruating 3 Place the patient in the proper position for the therapy 4 Use a sufficient amount of the appropriate contact medium 5 Set the prescribed intensity and time apply the emitter head and switch on the ultrasound 6 Move the emitter head continuously in accordance with the pre...

Страница 16: ...continuous ultrasound and at a max 1 0 W cm 2 for a pulse ultrasound with 1 2 PPR or greater Set the low frequency component intensity according to the object When searching for the trigger points set the threshold sensitive intensity outside the area where reflex changes are expected and with this intensity try to get as close as possible to the area The intensity will be above threshold sensitiv...

Страница 17: ...ad ERA The application time for dynamic method should be x times longer because the affected area is x times larger than the emitter head ERA An initial intensity of 0 4 0 6 W cm2 is used for continuous ultrasound 1 1 PPR For 1 2 1 4 PPR use the intensity 0 5 1 0 W cm2 and for 1 8 1 16 PPR use the initial intensity 1 0 1 3 W cm 2 The AMP intensity is chosen according to the effect The above thresh...

Страница 18: ...c in the injured area The simultaneous ultrasound application followed by increased capillary permeability improved extravasal liquid absorbtion increased vasoactive amines secretion and the influence on the pre capillar spfincters are considered the most effective prevention against the algodystrophic syndrom Physical therapy in the porouses area is contraindicated Use the segmental method only f...

Страница 19: ... subacuta 19 Epicondylitis acuta 20 Epicondylitis chronica 21 Epicondylitis subacuta 22 Haematoma chronica 23 Haematoma subacuta 24 Haematoma subchronica 25 Myalgia 26 Myorelaxatio 27 Myospasmus 28 Oedema chronica 29 Oedema subacuta 30 Oedema subchronica 31 Spondylarthrosis ancylopoetica M Bechterevi 32 Sy algodystrophicus Sy Sudeck 33 Tendinitis acuta 34 Tendinitis acuta subaqualis 35 Tendinitis ...

Страница 20: ...ion time 4 minutes daily 5x a week Total number of treatments 5 6 Apply to edema Patient must feel no pain during application 6 Bursitis acuta Emitter head 4 cm2 Frequency 1 MHz 1 8 PPR Intensity 1 2 W cm 2 Semistatic application Application time 1 3 minutes 0 5 minute steps daily 5x a week Total number of treatments 3 5 Apply to edema Pain must not be aggravated 7 Bursitis chronica Emitter head 4...

Страница 21: ...y 2 0 W cm 2 Application time 5 7 minutes 1 minute steps daily Total number of treatments 3 Subaqual application Emitter head distance 10 12 cm Water temperature 20 o C 17 Distorsio chronica Emitter head 4 cm 2 Frequency 3 MHz 1 1 PPR Intensity 1 0 1 5 W cm 2 Semistatic application Application time 5 minutes 3x a week Total number of treatments 6 9 Apply to chronic edema 18 Distorsio subacuta Emit...

Страница 22: ...eatments 5 8 In order to have long term effect physiotherapy must immediately follow irradiation 26 Myorelaxatio Emitter head 4 cm2 Frequency 1 MHz for superficial muscles 3 MHz 1 1 PPR Intensity 1 5 2 0 W cm 2 step 0 2 W cm2 Dynamic application Application time 5 minutes 3 5x a day Total number of treatments 5 8 In order to have long term effect physiotherapy must immediately follow irradiation 2...

Страница 23: ...step 0 2 W cm 2 Application time 4 minutes 5x a week Total number of treatments 2 5 Subaqual application Emitter head distance 10 12 cm Water temperature 18 25o C Treatment must include immobilization 35 Tendinitis chronica Emitter head 4 cm 2 Frequency 3 MHz 1 1 PPR Intensity 1 0 1 5 W cm 2 step 0 1 W cm 2 Semistatic application Application time 5 minutes 2 3x a week Total number of treatments 5 ...

Страница 24: ...d immediately follow irradiation Apply to the triggers that feature typical radiation when detected by either palpation or combined therapy 44 Trigger points superficialis Emitter head 1 cm2 Frequency 3 MHz 1 4 PPR Intensity 0 5 1 0 W cm 2 step 0 1 W cm 2 Semistatic application Application time 2 minutes per trigger 2x a day 5x a week Trigger points are severely sore points usually situated in the...

Страница 25: ...AM NUMBERS FOR COMBINED ULTRASOUND THERAPY ELECTROTHERAPY 51 Lumbalgie 52 Myalgia mm superficiales 53 Myalgia mm profundi 54 Relaxatio mm superficiales 55 Relaxatio mm profundi 56 Trigger points mm superficiales 57 Trigger points mm profundi 58 Zona hyperalgesica ...

Страница 26: ...tal number of treatments 1 6 55 Relaxatio mm profundi Emitter head 4 cm2 Frequency 1 MHz continual PPR Intensity 0 5 1 0 W cm 2 step 0 1 W cm 2 Electrotherapy AMP FM 100 200 Hz steps 3 3 program 0947 Contralateral electrode placement Intensity at threshold motor Application time 1 10 minutes step 1 minute daily Total number of treatments 1 6 56 Trigger points superficialis Emitter head 1 cm2 Frequ...

Страница 27: ...and maintain it at a threshold sensitive C Therapy effect control Set identical parameters as in A including the current intensity and move the head as in A If the therapy is successful the original TP will no longer exist Otherwise repeat the B procedure If a palpably or clinically diagnosed trigger cannot be detected by the method described in A repeat the whole procedure with the ultrasound int...

Страница 28: ... any means incl electronic mechanic photographic or other records without previous approval from BTL Industries Limited BTL Industries Limited operates a policy of continuous development Therefore it reserves the right to make changes and improvements to the Product described in this manual without prior notice The contents of this document is provided as is Except as required by applicable law no...

Отзывы: