BTL – 07p
operating manual & user's guide
page 22
Application time: 3 - 5 minutes, 1 minute steps, 3 - 5x a week. Total number of treatments: 6 - 9.
Apply to fibres with inner incoordination or to previously diagnosed trigger points.
Direct application of ultrasound to the epicondyle is contraindicated!
22. Haematoma chronica
Emitter head: 4 cm
2
. Frequency: 3 MHz, 1:1 PPR. Intensity: 1.0 - 1.2 W/cm
2
, step 0.1 W/cm
2
. Dynamic
application.
Application time: 4 minutes, 3x a week. Total number of treatments: 3 - 6.
23. Haematoma subacuta
Emitter head: 4 cm
2
. Frequency: 3 MHz, 1:4 PPR. Intensity: 1.0 - 2.0 W/cm
2
, step 0.1 W/cm
2
. Dynamic
application.
Application time: 4 minutes, 3 - 5x a week. Total number of treatments: 3 - 5.
24. Haematoma subchronica
Emitter head: 4 cm
2
. Frequency: 3 MHz, 1:2 PPR. Intensity: 1.0 - 2.0 W/cm
2
, step 0.1 W/cm
2
. Dynamic
application.
Application time: 4 minutes, 3 - 5x a week. Total number of treatments: 3 - 5.
25. Myalgia
Emitter head: 4 cm
2
. Frequency: 1 MHz (for superficial muscles 3 MHz), 1:1 PPR. Intensity: 1.5 - 2.0 W/cm
2
,
step 0.2 W/cm
2
. Dynamic application.
Application time: 5 minutes, 3 - 5x a day. Total number of treatments: 5 - 8.
In order to have long-term effect physiotherapy must immediately follow irradiation.
26. Myorelaxatio
Emitter head: 4 cm
2
. Frequency: 1 MHz (for superficial muscles: 3 MHz), 1:1 PPR. Intensity: 1.5 - 2.0 W/cm
2
, step
0.2 W/cm
2
. Dynamic application.
Application time: 5 minutes, 3 - 5x a day. Total number of treatments: 5 - 8.
In order to have long-term effect physiotherapy must immediately follow irradiation.
27. Myospasmus
Emitter head: 4 cm
2
. Frequency: 1 MHz (for superficial muscles: 3 MHz), 1:1 PPR. Intensity: 1.5 - 2.0 W/cm
2
, step
0.2 W/cm
2
. Dynamic application.
Application time: 5 minutes, 3 - 5x a day. Total number of treatments: 5 - 8.
In order to have long-term effect physiotherapy must immediately follow irradiation.
28. Oedema chronica
Emitter head: 4 cm
2
. Frequency: 3 MHz, 1:1 PPR. Intensity: 1.0 - 1.2 W/cm
2
, step 0.1 W/cm
2
. Dynamic
application.
Application time: 4 minutes, 3x a week. Total number of treatments: 3 - 6.
29. Oedema subacuta
Emitter head: 4 cm
2
. Frequency: 3 MHz, 1:4 PPR. Intensity: 1.0 - 1.2 W/cm
2
, step 0.1 W/cm
2
. Dynamic
application.
Application time: 4 minutes, 3 - 5x a week. Total number of treatments: 3 - 5.
30. Oedema subchronica
Emitter head: 4 cm
2
. Frequency: 3 MHz, 1:2 PPR. Intensity: 1.0 - 1.2 W/cm
2
, step 0.1 W/cm
2
. Dynamic
application.
Application time: 4 minutes, 3 - 5x a week. Total number of treatments: 3 - 5.