7. Troubleshooting
IntelliXcap
Error Messages
Part Number: 319430 Rev. A
Error
Code
Cause
Corrective Action
E149
Timeout Open Tray operation in recovery
mode.
Call service.
E150
Timeout homing cartridge sequence in
Store command.
Power cycle instrument.
Call service.
E151
Timeout Close Tray operation in recovery
mode.
Call service.
E152
Decap Error detected in decap retry
operation.
See
Remove any partial screwed on caps.
Initialize the instrument.
Retry the operation.
Call service.
E153
Closing tray sequence timeout in
recovery mode close tray operation.
Call service.
E154
Timeout Close Tray sequence in close
tray operation.
Call service.
E155
Timeout opening stage sequence in
Open Tray operation.
Call service.
E156
Main Z operation timeout in init with caps
operation.
Call service.
E157
Timeout stage sequence in Init with caps
operation.
Call service.
E158
Timeout in cartridge homing sequence in
recap operation.
Call service.
E159
Timeout in stage homing sequence in
recap operation.
Call service.
E160
Timeout opening safety door sequence in
Standby operation.
Verify there are no objects blocking the safety
door.
Power cycle the unit.
Run a Standby operation.
Call service.
E161
Main Z operation timeout in init with caps
operation.
Call service.
E162
Timeout stage/nest sequence in recovery
mode.
Call service.
37
Copyright © 2019, Brooks Automation, Inc.