Brooks Automation
3. Specifications and Site Requirements
Part Number: 319430 Rev. A
Site Requirements
Environmental Requirements
The IntelliXcap shall be used within the rule set of the Good Laboratory Practices, GLP.
The machine must be operated indoors and under the following environmental specifications only:
Parameter
Specification
Temperature - Transport and
Storage
15 to 40 ºC (59 - 104 ºF)
Temperature - Operation
0 to 40 ºC (32 – 104 ºF) Using the IntelliXcap in an environment
where the temperature is 40 ºC (104 ºF) or higher for an extended
period may cause the screen contrast level of the monochrome LCD
to decrease from its original level of brightness.
Storage Humidity
10 to 70% RH
Wet bulb temperature 39 ºC (102 ºF) max., no condensation
Relative Humidity
10 to 90% RH Wet bulb temperature 39 ºC (102 ºF) max., no
condensation
Storage Lighting
All external surfaces are resistant to UV-light. Over time UV-light
might affect LCD-panels: LCD screens may fade.
IP 30
Protection against small foreign bodies > 2.5 mm (e.g. a screwdriver),
and no protection against water
Dust
0.1 mg/m
3
and below (non-conductive levels)
Pollution Degree
For use in Pollution Degree 2 environment
Decontamination treatment with Hydrogen Peroxide Gas needs to be
avoided as it will damage the electronic parts.
Table 3-3: Environmental Requirements
Copyright © 2019, Brooks Automation, Inc.
14