6. Preventative Maintenance
IntelliXcap
Schedules and Procedures
Part Number: 319430 Rev. A
Unauthorized Service
Personal injury or damage to equipment may result if this product is operated or
serviced by unauthorized personnel.
l
Only qualified personnel are allowed to transport, assemble, operate, or maintain
the Product.
l
Properly qualified personnel are those who have received certified training and
have the proper qualifications for their jobs.
Parts
Brooks Automation can provide all parts required for preventive maintenance. For a list of these
parts, contact Brooks Automation Technical Support. To obtain additional information about parts
for preventative maintenance, contact your local Brooks Sales Representative, or call Brooks
Automation Technical Support. See
.
Schedules and Procedures
The service life of the IntelliXcap is 20 years of a daily 8-hour operation. This is based on the
presumption that all service and maintenance instructions described in this instruction manual are
observed.
Maintenance Schedule
Servicing the machine must only be carried out by qualified personnel. Tasks may require skills
and training. These instructions are a minimum requirement and must be carried out according to
the plan below.
Keep a logbook, or similar, to document the maintenance and cleaning schedules.
Task
Recommended Service Interval
Cap-Driver
Cartridge
IntelliXcap
General Visual
Inspection
2,500 cycles
NA
Preventative
Maintenance Visit
NA
20,000 cycles or 12 months which ever comes
sooner
Exchange
5,000 cycles
At 40,000 cycles it is recommended that the Head
is replaced
Table 6-1: Preventative Maintenance Schedule