Section 2 Installation
Brooks
®
Ar-Mite
TM
MT3750C
19
For electrical hook-up of units requiring intrinsically safe power
connections, use Brooks optional power supply and relay connected as
in Figure 2-6 for the integral alarm.
Installation of Transmitter
A. For intrinsically safe operation (recognized by a BLUE COVER),
intrinsic safety barrier selection, cable parameters and power supply
limits must be in accordance with the entity parameters shown in
Figure 2-6. Do not connect the zero wire (keep it floating) when the
transmitter is in normal operation. The zero wire is not meant to be
extended as zeroing must be done locally.
B. If the area classification is Division 2 and Zone 2, a barrier is not
required and cable parameters are not applicable. The electrical code
will require the use of conduit for wire protection. Refer to Figure 2-7.
C. If the installation is to be protected by Division 1 and Zone 1 explosion-
proof methods, explosion-proof installation methods must be followed.
Refer to Figure 2-8.
D. The power supply voltage determines the maximum resistance of the
loop, the associated cable and the barrier. Always be sure that the
voltage on the transmitter is within the specifications.
E. After installation and powering of the loop, the transmitter must be
zeroed to compensate for any stray magnetic effects in the vicinity
of the transmitter. The zero function may be activated as part of a
periodical maintenance check. The zero function is activated by
momentary (>2 seconds) shorting the zero wire to the ground wire
while power is on. Connect the transmitter as shown in Figure 2-9.
Содержание Ar-Mite MT3750
Страница 15: ...Section 1 Introduction 11 Figure 1 10 Dimensions for MT3750C Panel Mounting Brooks Ar MiteTM MT3750C...
Страница 19: ...Section 2 Installation 15 Figure 2 3 Inductive Switch Wiring Brooks Ar MiteTM MT3750C...
Страница 24: ...Section 2 Installation 20 Figure 2 7 Transmitter Wiring Diagram for I S Systems Brooks Ar MiteTM MT3750C...
Страница 25: ...Section 2 Installation 21 Figure 2 8 Transmitter Wiring Diagram for XP Systems Brooks Ar MiteTM MT3750C...
Страница 26: ...Section 2 Installation Brooks Ar MiteTM MT3750C 22 Figure 2 9 Transmitter Wiring Diagram...
Страница 30: ...Section 3 Operation Brooks Ar MiteTM MT3750C 26 Figure 3 1 Reed Switch Alarm...
Страница 34: ...Section 5 Parts List 30 Figure 5 1 Model MT3750 Exploded View Brooks Ar MiteTM MT3750C...