ESD (Electrostatic Discharge)

 CAUTION: This instrument contains electronic components that are susceptible to damage by static electricity. Proper handling procedures must be observed

during the removal, installation or other handling of internal circuit boards or devices.

Handling Procedure:

1. Power to unit must be removed.

2. Personnel must be grounded, via a wrist strap or other safe, suitable means before any printed circuit card or other internal device is installed,

removed or adjusted.

3. Printed circuit cards must be transported in a conductive container. Boards must not be removed from protective enclosure until immediately before

installation. Removed boards must immediately be placed in protective container for transport, storage or return to factory.

Comments

This instrument is not unique in its content of ESD (electrostatic discharge) sensitive components. Most modern electronic designs contain components

that utilize metal oxide technology (NMOS, SMOS, etc.). Experience has proven that even small amounts of static electricity can damage or destroy these

devices. Damaged components, even though they appear to function properly, exhibit early failure.

Brooks Instrument designs, manufactures and tests its products to meet many national and international standards. These products must be properly

installed, operated and maintained to ensure they continue to operate within their normal specifications. The following instructions must be adhered to

and integrated into your safety program when installing, operating and maintaining Brooks Instrument products.

To ensure proper performance, use qualified personnel to install, operate, update, program and maintain the product.

Read all instructions prior to installing, operating and servicing the product. If this instruction manual is not the correct manual, please see back cover

for local sales office contact information. Save this instruction manual for future reference.

 WARNING: Do not operate this instrument in excess of the specifications listed in the Instruction and Operation Manual. Failure to heed

this warning can result in serious personal injury and / or damage to the equipment.

If you do not understand any of the instructions, contact your Brooks Instrument representative for clarification.

Follow all warnings, cautions and instructions marked on and supplied with the product.

 WARNING: Prior to installation ensure this instrument has the required approval ratings to meet local and national codes. Failure to heed this warning can

result in serious personal injury and / or damage to the equipment.

Install your equipment as specified in the installation instructions of the appropriate instruction manual and per applicable local and national codes.

Connect all products to the proper electrical and pressure sources.

Operation: (1) Slowly initiate flow into the system. Open process valves slowly to avoid flow surges. (2) Check for leaks around the flow meter inlet

and outlet connections. If no leaks are present, bring the system up to the operating pressure.

Please make sure that the process line pressure is removed prior to service. When replacement parts are required, ensure that qualified people use

replacement parts specified by Brooks Instrument. Unauthorized parts and procedures can affect the product's performance and place the safe

operation of your process at risk. Look-alike substitutions may result in fire, electrical hazards or improper operation.

Ensure that all equipment doors are closed and protective covers are in place to prevent electrical shock and personal injury, except when

maintenance is being performed by qualified persons.

 WARNING: For liquid flow devices, if the inlet and outlet valves adjacent to the devices are to be closed for any reason, the devices must be completely

drained. Failure to do so may result in thermal expansion of the liquid that can rupture the device and may cause personal injury.

All pressure equipment with an internal pressure greater than 0.5 bar (g) and a size larger than 25mm or 1" (inch) falls under the Pressure Equipment Directive (PED).

The Specifications Section of this manual contains instructions related to the PED directive.

Products described in this manual are in compliance with EN directive 2014/34/EU.

All Brooks Instrument Flowmeters fall under fluid group 1.

Products larger than 25mm or 1" (inch) are in compliance with PED category I, II or III.

Products of 25mm or 1" (inch) or smaller are Sound Engineering Practice (SEP).

The Brooks Instrument (electric/electronic) equipment bearing the CE mark has been successfully tested to the regulations of the Electro Magnetic

Compatibility (EMC directive 2014/30/EU).

Special attention however is required when selecting the signal cable to be used with CE marked equipment.

Quality of the signal cable, cable glands and connectors:

Brooks Instrument supplies high quality cable(s) which meets the specifications for CE certification.

If you provide your own signal cable you should use a cable which is overall completely screened with a 100% shield.

“D” or “Circular” type connectors used should be shielded with a metal shield. If applicable, metal cable glands must be used providing cable screen clamping.

The cable screen should be connected to the metal shell or gland and shielded at both ends over 360 Degrees.

The shield should be terminated to an earth ground.

Card Edge Connectors are standard non-metallic. The cables used must be screened with 100% shield to comply with CE certification.

The shield should be terminated to an earth ground.

For pin configuration : Please refer to the enclosed Instruction Manual.

European Pressure Equipment Directive (PED)

European Electromagnetic Compatibility (EMC)

Essential  Instructions

Read before proceeding!

Содержание Ar-Mite MT3750

Страница 1: ...3750C Section 1 Introduction Ar MiteTM Low Flow Armored Flowmeter Installation Operation Manual Ar MiteTM Model MT3750 Metal Tube Flowmeter Ar MiteTM Model MT3750 Metal Tube Flowmeter with Transmitter...

Страница 2: ...r electrical and pressure sources Operation 1 Slowly initiate flow into the system Open process valves slowly to avoid flow surges 2 Check for leaks around the flow meter inlet and outlet connections...

Страница 3: ...3 Installation of Inductive Alarm 13 Installation of Reed Switch Alarm 16 Installation of Transmitter 19 Operation Section 3 Operating Procedure 23 Operation of Inductive Alarm 23 Operation of Transmi...

Страница 4: ...ram 14 2 3 Inductive Switch Wiring 15 2 4 Reed Switch Alarm Wiring Using IS Barriers 16 2 5 Reed Switch Alarm Installation Using P F Relay Unit 17 2 6 Reed Switch Alarm Wiring Using P F Relay Unit 17...

Страница 5: ...pressure and difficult to handle fluids Optional accessories include 4 20 mA output Needle Valve Flow Controllers and Alarms Specifications Pressure Equipment Directive PED 2014 08 EU Flowmeters metn...

Страница 6: ...tches Transmitter 4 20 mA output Agency Approvals Refer to Tables 1 6 thru 1 9 Optional Equipment Cartridge or NRSTM valves Integrally mounted flow controllers Meter Meter Meter Meter Meter Flow Range...

Страница 7: ...ion SEP CRN CRN IP66 67 NEMA 4X Declaration IP66 67 DEKRA Certificate UL IP65 DEKRA Certificate Type 4X CSA UL Certificate MBID 022 Explosion safety Constructional safety c ATEX EU Declaration of Conf...

Страница 8: ...9 Class I II III Div 1 Groups A thru G T6 Class I Zone 1 AEx ia IIC T6 Ex ia IIC T6 Ex ia IIC T6 Gb Ex iaD 21 T70 C GB3836 1 4 2010 GB 12476 4 2010 Ex nA II T6 Class I Div 2 Grps A B C and D Class II...

Страница 9: ...Section 1 Introduction Brooks Ar MiteTM MT3750C 5 Figure 1 3 Transmitter Wiring Diagram Figure 1 2 Power Supply vs Maximum Load Resistance...

Страница 10: ...losion safety Intrinsic Safety ia 1788748 Explosion safety Non incendive 1788748 Explosion safety Intrinsic Safety ia Class I Div 1 Groups A B C and D Class II Groups E F and G Class III Encl Type 4X...

Страница 11: ...C With Inductive Alarm Figure 1 6 Model MT3750C with Inductive Alarm Mechanical Transmitter Inductive Switch Alarm Reed Switch Alarm Explosion safety Intrinsic Safety ia ATEX KEMA 02ATEX1126 IECEX IEC...

Страница 12: ...Section 1 Introduction Brooks Ar MiteTM MT3750C 8 Figure 1 7 Dimensions for MT3750C Threaded Connections Metal Tube Flowmeter with Indicator...

Страница 13: ...Section 1 Introduction Brooks Ar MiteTM MT3750C 9 Figure 1 8 Dimensions for MT3750C Threaded Connections with Transmitter or Inductive Alarm...

Страница 14: ...Section 1 Introduction 10 Figure 1 9 Dimensions for MT3750C Threaded Connections with Reed Switch Alarm Brooks Ar MiteTM MT3750C...

Страница 15: ...Section 1 Introduction 11 Figure 1 10 Dimensions for MT3750C Panel Mounting Brooks Ar MiteTM MT3750C...

Страница 16: ...acking materials Inspect for damaged or missing parts Recommended Storage Practice If intermediate or long term storage of equipment is required it is recommended that the equipment be stored in accor...

Страница 17: ...s A Carefully remove the covers from each end of the flowmeter B Install the flowmeter with the inlet at the bottom and the outlet at the top C When installing the flowmeter in the process line follow...

Страница 18: ...y parameters shown on drawing B If the area classification is a Division 2 or Zone 2 a barrier is not required Note however when intrinsic safe apparatus is used without barrier this apparatus may NOT...

Страница 19: ...Section 2 Installation 15 Figure 2 3 Inductive Switch Wiring Brooks Ar MiteTM MT3750C...

Страница 20: ...m printed circuit board is a three terminal block marked Alarm Circuit The middle terminal is marked Alarm In This terminal is common for both high alarm and low alarm applications A circuit is comple...

Страница 21: ...Section 2 Installation Brooks Ar MiteTM MT3750C 17 Figure 2 4 Reed Switch Alarm Installation Using IS Barriers Ref No 981Z985 Rev G pg 1...

Страница 22: ...eTM MT3750C 18 Figure 2 5 Reed Switch Alarm Installation Using P F Relay Unit Figure 2 6 Reed Switch Alarm Wiring Using P F Relay Unit Note CSA US and Cananda Approved installations refer to Figures 2...

Страница 23: ...t required and cable parameters are not applicable The electrical code will require the use of conduit for wire protection Refer to Figure 2 7 C If the installation is to be protected by Division 1 an...

Страница 24: ...Section 2 Installation 20 Figure 2 7 Transmitter Wiring Diagram for I S Systems Brooks Ar MiteTM MT3750C...

Страница 25: ...Section 2 Installation 21 Figure 2 8 Transmitter Wiring Diagram for XP Systems Brooks Ar MiteTM MT3750C...

Страница 26: ...Section 2 Installation Brooks Ar MiteTM MT3750C 22 Figure 2 9 Transmitter Wiring Diagram...

Страница 27: ...he desired alarm flow rate E Move switch assembly until the metal initiator attached to the pointer is not inserted into the switch F Tighten the screw mounting the switch to the plate G Replace the i...

Страница 28: ...Flow 40 70 Follow the row across to the column marked Jumpers In this case the jumpers required are J1 J4 and J6 Simply pull the jumpers out from their current positions and place them in the correct...

Страница 29: ...flow alarm within a limited range of the maximum flow rate of the meter The dual setpoint alarm is set in the same manner as described in Section 3 4 Simply use two ohmmeters to observe the behavior...

Страница 30: ...Section 3 Operation Brooks Ar MiteTM MT3750C 26 Figure 3 1 Reed Switch Alarm...

Страница 31: ...Section 4 Maintenance Brooks Ar MiteTM MT3750C Section 4 Maintenance 27 General It is important that this device only be serviced by properly trained and qualified personel CAUTION...

Страница 32: ...and alone device consult factory for more details PC interface is needed to calibrate the transmitter With this interface the transmitter can be connected to the serial RS232 port of a computer At pow...

Страница 33: ...n 5 Parts List Section 5 Parts List 29 General When ordering parts please specify Brooks Serial Number Model Number Part Description Quantity Refer to Figure 5 1 and Tables 5 1 5 2 Brooks Ar MiteTM MT...

Страница 34: ...Section 5 Parts List 30 Figure 5 1 Model MT3750 Exploded View Brooks Ar MiteTM MT3750C...

Страница 35: ...SERVICE AND IN SITU CALIBRATION Brooks Instrument can provide start up service prior to operation when required For some process applications where ISO 9001 Quality Certification is important it is ma...

Отзывы: