Brookfield Engineering Labs., Inc.
Page 42
Manual No. M13-167-A0415
Table
: Display all data points. Data will be shown in a scroll list where each page can hold a
maximum of 50 points. If the data set has more than 50 points, then additional pages
will be indicated at the bottom of the screen. Additional pages of data can be accessed
through the use of the Blue Arrows
Each data point includes: Viscosity, Torque, Speed, Temperature, Time, Shear Stress
(SS), Shear Rate (SR), Density, and Accuracy.
Post Test Averaging
: Average and Standard Deviation are calculated for measured and calculated
parameters including: Viscosity, Torque, Shear Stress, and Temperature.
Post Test Averaging is calculated regardless of the Data Collection setting.
If Data Collection was set to Multipoint Averaging, then the Post Test
Averaging will calculate an average of the averaged data.
When using multiple step programs, the test step utilized for Post Test
Averaging is specified in the Report setting (see Section III.9).
Test Used
: Display the test elements used to generate the data set. In this view, the Configure
Test button is available. Selecting Configure Test will program the DV3T to run
the same test utilized to collect the data set being viewed and present the Configure
Viscosity Test screen.
Notes
: Document any relevant information about the test or data. This information will be
stored with the data set once saved.
Compare Results
: Load two data sets and view side by side.
Figure III-17
Device Info
: Display basic information about the data file and the specific DV3T Rheometer
used to collect the data including: date and time the test started, completed and
was saved; if user accounts are active, which user saved the data; rheometer serial
number, firmware version, rheometer Torque range (LV, RV, HA, HB, other).