Brookfield Engineering Laboratories, Inc.
Page 23
Manual No. M/98-211-B0104
Brookfield Engineering Laboratories, Inc.
Page 23
Manual No. M98-211-E0912
SPECIAL SPINDLE #2
NAME = D BB E
SMC = 1.000
SRC = 2.000
Figure II-14
To complete, enter 0 values for SMC and SRC.
The display may ask “ADD A SPINDLE”. You may press YES to repeat or NO to finish.
ADD A SPINDLE?
1 = YES
3 = NO
Figure II-15
If you press “3” NO, the display will ask “SAVE NEW SPINDLE?” Press the “1” YES to save or
press “3” NO.
SAVE NEW SPINDLES?
1 = YES
3 = NO
Figure II-16
The display will now say “SET UP COMPLETE”, turn power off.
You can edit any existing spindle entry names and spindle constants. If so, you will need to complete
the steps until you are prompted to “Save New Spindle”. For example, if you have three special
spindles entered and need to modify the second spindle, you will need to proceed through the third
spindle parameters by pressing “Enter” until prompted to “Save New Spindle”. This will then save
the modified spindle.
II.4 Direct Speed Entry
At this point, the user may choose to enter a speed by the so-called direct speed entry method.
Enter a valid speed in the range of
0.01
to
250
RPM
by pressing the numeric keys successively.
The previously blank line 3 on the default display screen records the user’s new speed input as
depicted in
Figure
II-17:
RPM: 0.0 SPINDLE: 31
TEMP: 72.1°F PRTN
ENTER NEW RPM: 11_
TORQUE = 0.0 %
Figure II-17
Here, the user intends to enter a speed of
112 RPM
, has pressed the “
1
” key twice and is about
to press the “
2
” key. If the user makes more than five (
5
) key presses, the
DV-III Ultra
control
program will “
roll
” the cursor back to the first character of the field and begin to overwrite the
previous data entry.
Next the user presses the
ENTER
key to accept the speed. The motor will begin running at
112 RPM