Viscometer Will Not Return To Zero
✓
Rheometer is not level
• Adjust the laboratory stand
✓
Pivot point or jewel bearing faulty
• Perform an Oscillation Check*
✓
Manually lift the spindle nut and rotate counter clockwise until the display shows
20% - 30% deflection.
✓
Release the spindle nut.
✓
The % display should return to 0.0%
±
0.1%.
• Perform calibration check
• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.
* This procedure does not apply to instruments with ball bearing suspension (See Section I.7).
❑
Display Reading Will Not Stabilize
✓
Special characteristic of sample fluid. There is no problem with the rheometer.
• Refer to Appendix C
✓
Check for erratic spindle rotation
• Verify power supply
• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.
✓
Bent spindle or spindle coupling.
• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.
✓
Temperature fluctuation in sample fluid
• Use temperature bath control.
❑
No Recorder Response
✓
Be sure the rheometer is not at ZERO reading.
✓
Be sure the recorder is ON and not on STANDBY.
✓
Verify the range settings.
✓
Check cable leads for clean connection.
❑
Recorder Pen Moves In Wrong Direction
✓
Output polarity reversed
• Reverse leads