Bronkhorst®
Instruction Manual MI Series Industrial mini CORI-FLOW™
9.17.120A
24
3.7.2.2
Power-up functions
·
In order to access these functions, press and hold the switch while powering up the instrument.
·
The available functions are presented in a repeating sequence of patterns, where each pattern indicates a function.
·
Indications in this sequence are flashing (0.2 sec on, 0.2 sec off).
·
To start a function, release the switch when the LEDs show the pattern of the required function.
•
(green)
•
(red)
Time
Function
•
•
0…4 sec
No action
•
•
4…8 sec
Restore factory settings
(except communication settings)
•
•
8…12 sec FLOW-BUS:
Auto install to bus
; let the instrument obtain a free node address from
the FLOW-BUS system
•
•
12…16 sec
Activate configuration mode
:
·
The baud rate and bus type for the Main terminal are set to 38400 and RS232
FLOW-BUS (ProPar) respectively
·
In configuration mode, the green LED blinks 2 seconds on and 0.1 second off
·
Configuration mode remains active after powering-down and can be deactivated
by selecting this function again at the next start-up
3.7.2.3
Control mode - readout/change
Reading control mode
·
By briefly pressing the switch 2 times with intervals of up to 1 second in normal operation mode, the instrument shows
its current control mode with a series of consecutive LED indication patterns.
·
The number of flashes corresponds to the current value of parameter
Control Mode
(see
Special parameters
).
Step Pattern
Indication
1.
Green
•
•
number of flashes indicates the tens of the parameter value
2.
Red
•
•
number of flashes indicates the units of the parameter value
Examples:
·
for value 1 (control mode 'Analog input'), the green LED will flash 0 times and the red LED 1 time
·
for value 22 (control mode 'Valve Safe State'), the green and red LED will each flash 2 times
Changing control mode
·
By briefly pressing the switch 4 times with intervals of up to 1 second in normal operation mode, the instrument enters a
state in which the control mode can be changed.
·
This is done in 2 steps, each represented by a LED indication pattern (green or red; see table below).
·
The number of flashes corresponds to the available values of parameter
Control Mode
(see
Special parameters
).
·
At the start of each step, the according LEDs starts flashing fast (0.1 second on, 0.1 second off). By pressing and holding
the switch, the associated action is started and the flashing slows (05. seconds on, 0.5 seconds off).
Step Pattern
Maximum
flash
count
Action
1.
Green
•
•
2
set tens of parameter value
2.
Red
•
•
9
set units of parameter value