The colors are based on the NAMUR NE 107 standard and are a simplified representation of the instrument's
.
·
After powering up or restarting the instrument, the indication light cycles through all status colors once.
·
After the startup cycle, the indication lights assumes the color associated with the current instrument status.
Bluetooth
·
When Bluetooth is active, the indication light flashes blue once every 3 seconds.
·
If the instrument has a Bluetooth connection, the indication light flashes blue twice every 3 seconds.
·
Bluetooth communication is not reflected by the indication light.
·
Bluetooth indications are not related to the blue NAMUR indication.
Zeroing
·
During
, the indication light blinks blue (1 second on, 1 second off).
·
The zeroing indication is not related to the blue NAMUR indication.
3.2
Special procedures
3.2.1
Address selection
Manual address selection
Assuming the default baud rate and parity match the fieldbus settings, the fieldbus address can be quickly set using the
manual
:
·
With value 0, the instrument uses the digital address setting
·
Values from 1 to 9 override the digital setting
Digital network settings
FlowSuite can be used to set the fieldbus address, baud rate and parity digitally (through the support interface or with
Bluetooth). This can even be done before connecting the instrument to the fieldbus.
See also parameter section
for digital parameter settings and default values.
·
A USB connection on the support interface provides sufficient power for instrument configuration only.
·
Bluetooth configuration requires powering the instrument through the power connector or support interface.
3.2.2
Switching control function
Communication channels
The so called multi-parameter functionality enables simultaneous measurement of flow, inlet pressure and outlet pressure,
using 3 different communication channels:
Communication
channel
Measurement and control variable
1
Flow (F)
2
Inlet/upstream pressure (P1)
3
Outlet/downstream pressure (P2)
17
16
7
27