Liquid Dosing Control Manual
9.17.150 rev A – September 2020
Page
25
of
34
Reset Batch Counter to start next batch
- Write value 3 to parameter 114 – Reset
(Reset counter)
Step DDE ProP
ar
Name
Data type Execute Value Remark
1
130
104/
8
Counter Mode
Uint8
write
2
Counter up to limit [batch mode]
2
9
or
205
1/1
33/3
Setpoint
Fsetpoint
Uint16
Float
write
x
x.x
Flow Setpoint integer 0-32000
Flow Setpoint Real value x.xx
3
114
115/
8
Reset
Uint8
write
3
Reset counter & Start batch
4.2 Pause dosing
Pause dosing
- Write value 3 to parameter 12 – Control Mode
(Actuator 0%)
Resume dosing
- Write value 0 to parameter 12 – Control Mode
(Restore to previous)
In case of Batch Dosing, no action is required to Pause the dosing
Step DDE ProPar Name
Data
Type
Execute Value
Remark
Pause Dosing
1
12
1/4
Control Mode
Uint8 Write 3
Actuator 0%
Resume Dosing
2
12
1/4
Control Mode
Uint8
Write 0
Digital setpoint (bus)
4.3 Stop dosing
Stop controller
- Write value 0 to parameter 206 – Flow Setpoint
(Setpoint)
or
- Write value 0 to parameter 9 – Setpoint
(Setpoint)
Extra step for Batch Dosing
- Write value 0 to parameter 130 – Counter Mode
(Counte Mode OFF)
Check hardware
-
Ensure all valves in the system are set correctly to prevent unwanted flow