Liquid Dosing Control Manual
9.17.150 rev A – September 2020
Page
23
of
34
4. Operate the Liquid Dosing Set
4.1 Start dosing
Check hardware
- Ensure all valves in the system are set correctly for operation
4.1.1 Continuous Dosing
Provide flow setpoint to controller
- Write value 0 to parameter 12 – Control Mode
(Setpoint RS232 / BUS)
This configures the instrument to receive a setpoint via the communication interface
(RS232 or Fieldbus). The Field bus will overrule any changes made via RS232.
or
- Write value <___> to parameter 206 – Flow Setpoint
(Setpoint)
Value to be defined in the previously selected capacity unit
(kg/h, gr/h, L/min, etc..)
or
- Write value <___> to parameter 9 – Setpoint
(Setpoint)
Value based on flow percentage of Full Scale flow
Value 0 = 0%
Value 32000 = 100%
Step DDE ProPar Name
Data
Type Execute Value
Remark
Start Dosing
1
12
1/4
Control Mode
Uint8
Write 0
Digital setpoint (bus)
1
206
33/3
fSetpoint
Float write
<value> In Capacity Unit.
1
9
1/1
Setpoint
Uint16 write
<value> 0-32000 = 0-100%