3.3 Verify system status
The system readiness for operation can be checked by monitoring the below parameters
(assuming Idle condition)
Internal temperature sensor
When the internal temperature sensor provides a stable measurement value, then this is an
indication of a stable liquid supply temperature, stable ambient temperatures, and a completion of
the warm-up time of the instrument.
- Read parameter 142 – Temperature
(Temperature in °C)
Please note that upstream temperature changes are not detected by the instrument when there is
no flow. However, the instrument itself can warm up causing a rise in temperature.
Density (Coriolis sensor)
The density measurement provides an indication of any gas trapped in the liquid (bubbles) or a
change in the concentration or composition of the fluid. This measurement is affected by
temperature changes.
- Read parameter 270 – Actual Density
(Density in kg/m
3
)
Example
H
2
O at room temperature
1001.91943359375 kg/m
3
Zero stability
Process or application specific conditions can influence the zero stability. Factors include ambient
temperatures, gas dissolved in the fluid, or switching fluids.
- Read parameter 8 – Measure
(Measure 0 – 32000)
Value 0 corresponds with 0% of the Capacity Value (Full Scale)
Value 32000 corresponds with 100% of the Capacity Value (Full Scale)
or
- Read parameter 205 – fMeasure
(Measure flow)
Value is measurement value in selected Capacity Unit
Please note that small pressure changes up- or downstream can create a flow through the sensor
and can therefore cause an unstable zero-stability (flow measured when no flow should be
present). For calibrating an accurate zero stability it is crucial to prevent this type of flow by
installing shut-off valves.
A pump does not function as a closed valve when the setpoint is 0.
In the next section, the process to execute an Auto-Zero function is described.
Step DDE ProPar Name
Data
Type Execute Value
Remark
1
142
33/7
Temperature
Float read
<temperature> In degrees Celcius
2
270
116/15 Actual Density
Float read
Measured density [kg/m3]
3
8
1/0
Measure
Uint16 read
<measure>
0-32000 = 0-100%
4
205
33/0
fMeasure
Float read
<measure>
In Capacity Unit.