background image

 

 

Leaving the patient unattended in the chair near other objects. 

 

Leaving a potentially agitated patient in an unsupervised area. 

 

Leaving a patient in a chair on a sloping surface. 

 

Leaving a chair unattended on a sloping surface. 

 

Leaving a chair unattended without setting caster locks. 

 

Transferring a patient into or out of the chair without setting caster locks. 

 

Using the BRODA chair at temperatures below 0° Celsius (32

°

 Fahrenheit). 

 

1.12 

Cleaning 

 
BRODA Chairs should be wiped clean with soap and water. Diluted household strength ammonia 
or chlorine based cleaner, as well as a hospital grade cleaner may be used if necessary.  
 
Refer to the following guide for cleaning certain staining agents (Applies to the Permablok3® 
vinyl, as per the vinyl manufacturer’s recommendations).  The frame and components of the chair 
can be cleaned using the same procedure as the vinyl.  The steps below are also located in the 
Cleaning Guide found on the back of the Broda chair. 
 

Staining Agent Cleaning Procedure 

 

Eye Shadow / Mascara   

 

 

 

 

Step 1  

Grease / Suntan Lotion / Chocolate  

 

 

 

Step 1  

Blood / Bodily Fluids  

 

 

 

 

 

Step 2  

Red Lipstick  

 

 

 

 

 

 

Step 2  

Oil Base Paint / Tar / Asphalt  

 

 

 

 

Step 3  

Other Tough Stains  

 

 

 

 

 

Step 4 or 5*  

 

Step 1: 

Remove excess spill with a damp cloth. Clean with a 1:1 mix of Ivory® liquid and water. 

Rinse with clean water and dry.  
 

Step 2: 

Use a straight application of concentrated cleaners such as Formula 409® or Fantastik® 

Spray Cleaner. Then wipe with a clean cloth. If using a hospital grade cleaner, follow the diluting 
instructions on the label.  
 

Step 3: 

Use a 1:1 mix of ammonia and water or a 1:4 mix of bleach and water. Rinse with clean 

water and dry.  
 

Step 4: 

Use a straight application of naphtha (lighter fluid). Rinse thoroughly with clean water and 

pat surface dry.  
 

Step 5: 

Use a 1:1 mix of isopropyl alcohol and water. If the stain persists, use straight alcohol. 

Rinse thoroughly with clean water and pat dry. If the stain remains, use a 1:1 mix of acetone and 
water. Rinse with clean water and pat surface dry.  
 

*Note: 

for cleaning that requires steps 4 or 5 – Use a soft cotton cloth saturated with the cleaning 

material and rub the stain in circles 10 times. Pat dry with another soft cotton cloth, and check 
results. 
 
Pay close attention to the frame, sockets, fasteners, parts and casters, as well as the padding for 
an even, thorough clean. 
 
Thoroughly clean all removable padding, as well as the frame and components around the parts 
and padding that have been removed. 
 
BRODA Chairs should not be cleaned with petroleum based cleaners. Any petroleum based 
products that come in contact with any vinyl surface should be removed as quickly as possible. 
Petroleum based products make vinyl brittle and will damage the seating surface and cushions. 

Содержание LT tilt

Страница 1: ...5 Operating Manual Distributor Rep This manual must be given to the caregiver responsible for this chair and its occupant Caregiver Before using this chair read this manual thoroughly and save for fut...

Страница 2: ...WARNING 8 1 9 Accidental Motion Danger of Falling or Collision WARNING 8 1 10 Improper Restraint Use Risk of Serious Injury WARNING 8 1 11 Improper Use WARNING 8 1 12 Cleaning 9 1 13 Preventative Mai...

Страница 3: ...est Angle Adjustable Independently Elevating Seat Back Surface BRODA Comfort Tension Seating Padding Removable Padding seat and back pad neckrest side panels legrest GENERAL INSTRUCTIONS The instructi...

Страница 4: ...ions that are included in this manual are very important to the safe and effective operation of the BRODA Chair Safety requirements that are detailed in Section 1 must be followed at all times BRODA a...

Страница 5: ...e professional Do not lean forward out of the chair any further than the length of the armrests Make sure the casters are pointing forward when leaning forward Otherwise the chair may tip over Do not...

Страница 6: ...ly both to the first use of the Chair and to all subsequent uses An in service on the operation and safety issues as described in this manual must be given to the patient s caregivers prior to its fir...

Страница 7: ...from pulling the chair over or pulling themselves off the seating surface and to prevent the patients from pulling moveable objects onto the chair and onto themselves We recommend that the chair be u...

Страница 8: ...and exterior doorways Placing the patient near handrails tables or stationary objects while the caster locks are in place should be done with caution and should be supervised Failure to do so may lead...

Страница 9: ...id and water Rinse with clean water and dry Step 2 Use a straight application of concentrated cleaners such as Formula 409 or Fantastik Spray Cleaner Then wipe with a clean cloth If using a hospital g...

Страница 10: ...protected Leaving the chairs outdoors could cause the paint to peel Formula 409 is a trademark of the Clorox Company Fantastik is a trademark of DowBrands Inc Trademark of the Dow Chemical Company Ivo...

Страница 11: ...nance do not use any lubricants that contain solvents Solvents will damage many of the moving components on the chair If necessary a white food grade grease lubricant may be used on the sliding compon...

Страница 12: ...caregiver with the particular safety limits of the chair It is important that if assistance is required with the use of the chair that the caregiver remember to bend their knees and keep their back st...

Страница 13: ...seat Transfer s refers to the movement of a resident into or out of a chair with the assistance of their caregiver s Mechanical Transfer s refers to the movement of a patient into or out of a chair w...

Страница 14: ...upright condition and not stacked higher than 2 boxes tall No other materials should be shipped or stored on top of a BRODA box BRODA boxes should not be placed on pallets BRODA Chairs should be ship...

Страница 15: ...local BRODA representative who supplies the chair or the HME Provider The caregiver s must adhere to the Safety Requirements at all times to ensure the safety of the patient caregivers and third parti...

Страница 16: ...ting Seat Tilt 1 Squeeze the tilt handle which is located on the push bar The tilt will be easier to operate with a patient weight in the chair 2 Raise or lower the seat tilt until the desired angle i...

Страница 17: ...r comes to an abrupt stop Check to ensure casters are clean and free of dirt Check to ensure all caster locks are functioning Check for loosening of the attachment between caster stem and housing Chec...

Страница 18: ...functional testing believes that any function is not operating correctly the chair should be taken out of service until a satisfactory functional test can be completed The caregiver or maintenance de...

Страница 19: ...uator housing 1 Loosen the nut on the bottom of the cylinder Unclip the cable from the actuator lever on the cylinder 2 Cut the cable ties that secure the cable to the frame of the chair carefully not...

Страница 20: ...cated inside the barrel of the cylinder allowing for changes in position When the operator releases the handle the valve closes and locks the cylinder in the desired position The function of the actua...

Страница 21: ...rom the vhair to make this adjustment 1 Loosen the lock nut located on the stem of the gas cylinder next to the actuator with a 17mm wrench 2 From the above instructions determine whether you want to...

Страница 22: ...rts or warranty repairs The owner may return products to BRODA for warranty replacement or repair by shipping items prepaid and insured to the factory Warranty completed at the factory includes both m...

Отзывы: