138
PERFORMANCE SPECIFICATIONS
11
Range, resolution, and
accuracy of respiratory rate
monitor
Range: 1 BPM to 120 BPM
Resolution: 1 BPM
Accuracy: ±10% or 1 BPM, whichever is greater, up to 50 BPM
PARAMETER
DESCRIPTION
Range, resolution, and
accuracy of air or oxygen
minute volume display
Range: 0 LPM to 20 LPM.
Resolution: 0.1 LPM.
Accuracy: ±10% or ±(10 ml x average breath rate), whichever is greater.
NOTE:
The displayed air or oxygen minute volume is the product of the average
breath rate times the average measured air or oxygen volume settings.
During continuous delivery, the ventilator displays an air or oxygen minute volume of
3.0 LPM.
Oxygen source time
(stationary oxygen)
Indefinite
Usage types
Stationary use: place on table or flat surface.
Extended Range use with oxygen hose.
Stand-Alone use with cylinder or alternate gas source.
Ventilator Battery Specifications
Type
Lithium ion, rechargeable, not user replaceable.
Ampere/hour rating
1800 mAh
Maximum current
450 mA
Operating Voltage
7.4 V internal
Duration
Approximately 4 hours
If the battery performance degrades to unacceptable levels before the scheduled 2.5 year replacement, contact
your Breathe Technologies® service representative.
Compressor Battery Specifications
Type
Lithium ion, rechargeable, not user replaceable.
Ampere/hour rating
10200 mAh
Maximum current
10 A
Operating Voltage
28.8 V internal
Duration
1 hour
If the battery performance degrades to unacceptable levels before the scheduled 2.5 year replacement, contact
your Breathe Technologies® service representative.
Special Features
Alarm vibrate
Instead of an audio alarm, the user can enable the alarm to vibrate for medium- and
low-level alarms.
Touch screen flip
User can flip the screen orientation 180°.