ALARMS, ALERTS, AND TROUBLESHOOTING
SILENCING AND CLEARING
ON-SCREEN
ALARMS
Alarm notifications that appear on the touch screen originate from the ventilator. Silencing and clearing on-screen
alarms is a multi-step process that depends on alarm priority and how many alarms are active.
Press the Silence Alarm button to temporarily silence the on-screen alarm for 60 seconds. Pressing the
Silence Alarm button silences only one alarm at a time—in audible or vibrating alarm mode. If more than
one on-screen
alarm occurs, press the Silence Alarm button once for each alarm. If the on-screen
alarm is a
medium- or high-priority alarm and is not silenced after 60 seconds, the alarm will continue with an additional
buzzer.
Resolve the condition that triggered the on-screen alarm. For help resolving alarms, see the alarm tables
that follow for possible causes of an alarm and options to resolve it. If a Silence Alarm button is pressed but
the condition that triggered the alarm is not resolved, the alarm will sound again after 60 seconds.
After resolving a High Temperature, High Circuit Pressure, or High PEEP
Pressure high-priority alarm, touch
OK
in the message that indicates the
alarm has been resolved.
500 ml
High Circuit
Peak Pressure
has been resolved.
10:11a, Fri
Jun 12, 2015
3
Silence Alarm button on ventilator (when in Extended
Range or Stand-Alone Configuration).
Silence Alarm button on compressor (when in
Stationary Configuration).
Values displayed on the screen are for
illustrative purposes only.