120
MAINTENANCE
8
TESTING SETUP
Undock the ventilator from the compressor. For more information, see "Undocking the Ventilator from the
Compressor" on page 41.
Power on the ventilator. If necessary, record clinical settings before
changing settings.
Using the Clinician’s Menu, activate only the
Low Activity Prescription
Setting
.
Set the
Ventilation Settings
to:
Ventilation Settings
10:11a, Fri
Jun 12, 2015
ml
FLIP
CANCEL
OK
Volume
300
I-Time
PEEP
BR
sec
cmH
2
O
/min
0.75
0
12
0-9, OFF
Sensitivity
4
Touch
OK
and then
CONFIRM
.
Set the
Alarm Limits
to:
Set Alarm Limits
10:11a, Fri
Jun 12, 2015
/min
FLIP
CANCEL
OK
High BR
40
Low BR
High PIP
/min
cmH
2
O
0
40
cmH
2
O
Low PIP
1
Touch
OK
and then
CONFIRM
.
Set the
Breath Timeout
to:
Breath Timeout
20
Sec
12
BPM
10:11a, Fri
Jun 12, 2015
Touch
OK
and then
CONFIRM
.
Set the
Source Gas
to whatever source gas is being used for testing.
Connect the adjustable source gas supply to the
ventilator using an adjustable regulator set to 50
PSI. Turn on the source gas supply to 50 PSI
nominal.
Connect the interface. For more information, see
"Chapter 5: Connecting an Interface" on page 51.
NOTE:
The interface does not need to be worn during testing.
Place the ventilator on a flat surface where the interface connection will
not be jostled. Set the interface aside, with nasal pillows and tubing free
of anything that might impede air flow.
Prescription Settings
10:11a, Fri
Jun 12, 2015
3