129
BATTERY INFORMATION
The Life2000® Ventilation System uses lithium ion batteries with the following specifications.
VENTILATOR BATTERY SPECIFICATIONS
SPECIFICATION
DESCRIPTION
Type
Lithium ion, rechargeable, not user replaceable.
Ampere/hour rating
1800 mAh
Maximum current
450 mA
Operating Voltage
7.4 V internal
Duration
Approximately 4 hours
Ventilator
If the battery performance degrades to unacceptable levels before the scheduled 2.5
year replacement, contact your Breathe Technologies® service representative.
TIPS:
• A fully charged battery in good condition is designed to operate for four hours of typical use, but exact
operating time depends on patient breath rate.
• If the battery has no charge, it takes approximately three to four hours to fully recharge.
• The ventilator can be used while charging.
• The ventilator sounds an alarm and displays an alarm message when the battery has less than 25% of its
charge and then again when it has less than 15%.
CHAPTER 9: ACCESSORIES
COMPRESSOR BATTERY SPECIFICATIONS
SPECIFICATION
DESCRIPTION
Type
Lithium ion, rechargeable, not user replaceable.
Ampere/hour rating
10200 mAh
Maximum current
10 A
Operating Voltage
28.8 V internal
Duration
1 hour
Compressor
If the battery performance degrades to unacceptable levels, contact your Breathe
Technologies® service representative.
NOTE:
The compressor contains a lithium ion battery. Before shipping or traveling with the compressor, contact
your transportation carrier for information.
TIPS:
• A fully charged battery in good condition is designed to operate for one of typical use, but exact operating
time depends on usage conditions.
• If the battery has no charge, it takes approximately three to four hours to fully recharge.
• A fully charged battery in good condition lasts one hour.
• The compressor can be used while charging.
• The compressor sounds an alarm when the battery charge drops to 20% or less.