![Braun D521 series Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/braun/d521-series/d521-series_product-manual_2804885030.webp)
D521-Manual_EN_Rev07 2017_05_26
Page 30 of 31
8
Safety Data
8.1
Safety Data of Measuring Chain D521-A5S0 resp. D521-D461-A5S1
Note:
PFD values below are valid only if „Safety Notes for Installation and Operation“ are observed and maintained.
PFD values of D521 apply to relay output SP1 and analog output.
PFD Values at Tproof
Tproof
2 years
Tproof
5 years
Tproof
10 years
Tproof
20 years
A5S05…
A5S07..
A5S08..
A5S09..
A5S0DD0..
A5S0DS0..
A5S10
A5S11..
A5S12..
A5S13..
A5S1DD0..
A5S1DS0..
4.27 E-05
1.04 E-04
2.07 E-04
4.13 E-04
D461…
2.45 E-04
6.01 E-04
1.19 E-03
2.38 E-03
D521
…
2.52 E-04
6.29 E-04
1.26 E-03
2.51 E-03
Chain
D521-A5S0
2.53 E-04
6.30 E-03
1.27 E-03
2.52 E-03
Chain
D521-
D461-A5S1
4.97 E-04
1.23 E-03
2.45 E-03
4.90 E-03